7.3 Assembly and installation
The correct location is a key factor in the accuracy of the weighing results.
Levelling
Exact alignment and stable installation are an essential
requirement
for r
compensate for m inor unev enness or t ilts o f t he
footprint by levelling the weighing balance.
Level balance with foot screws until the air bubble of the
water balance is in the prescribed circle.
Check levelling regularly.
Installation of weighing plate
Connection to the power supply (USB)
using the enclosed power pack or the USB host (e.g. computer)
In or der t o obtain e xact r esults w ith t he el ectronic bal ances, t he di gital
weighing platform must have reached the operating temperature (see warming
up time chap. 1). The digital weighing platform must be connected to the power
supply during this warming up period.
The ac curacy o f t he di gital weighing pl atform i s de pendent o n t he l ocal
acceleration due to gravity.
Strictly observe hints in chapter Adjustment.
Connection to your system (configuration) as described in the chapter 8
Carry out adjustment as described in the chapter 9
Communication and using the balance via the interface protocol (see
interface protocol description)
The c hapter 10.1 describes t he i deal w ay t o us e t he balance us ing t he
BalanceConnection software.