Spectro LFP qb - Operating Manual

7.5 Warning and Error messages

The display of the Spectro LFP qb spectral unit can show the following
warnings/ errors:
W01: Wrong spectral unit detected. Please attach spectral unit with
serial number Cxxxxxx and restart the instrument.
The Spectro LFP qb Platform has stored parameters about the
spectral unit. If a new/different spectral unit is connected, a warn-
ing appears. Either connect the correct spectral unit or contact
support for receiving a new parameter file for the platform.
W02: Battery not inserted or bad battery! Insert battery or contact sup-
port if battery is defective.
The Spectro LFP qb can only be operated if the battery pack
is correctly inserted into the spectral unit. Please remove the
spectral unit from the Platform and verify if the battery pack
has been inserted
W03: Transport arm is not aligned. Please see instruction manual.
When transport arm is moving to its left position, two switches
indicate if the arm is correctly alligned. If it is misaligned (for
example after transport), this warning will show up.
Attach a sample holder to the
Spectro LFP qb and move the
Transport arm in its initial po-
sition. The edge of the sample
holder needs to be parallel to
the edge of Specrto LFP qb
platform. A distorted transport
arm looks like the following:
43