Appendix A
Specifications for analog output version

Analog output

Setpoint relays
Relay 1
Relay 2
Analog I/O male connector on module end panel
11–26 V power input 3
76
0.375 to 5.659 VDC for 0 to 1000 Torr, non-linear, absolute of N
Two single-pole, double-throw (SPDT) relays. Use potentiometers and test
points on the module front panel and jumpers inside the module to
configure relays.
• Relay 1 is assigned to the Convectron gauge, which indicates vacuum
chamber pressure
• Relay 1 activates with decreasing and deactivates with increasing
vacuum chamber pressure
• Hysteresis: 30 mV
• Range: 1 x 10
• Relay 2 is assigned to the differential pressure sensor, which indicates
pressure differential between atmospheric and vacuum chamber
pressures
• In default mode, relay 2 activates with increasing and deactivates with
decreasing differential between atmospheric and vacuum chamber
pressures
• Hysteresis: 8 Torr
• Range: –750 to +125 Torr
Analog output 5
Power ground 4
Setpoint 2 adjust 2
Setpoint 1 adjust 1
–3
to 1000 Torr
15 Relay 1 common
14 Relay 1 normally closed
13 Relay 2 normally closed
12 Relay 2 common
11 Relay 2 normally open
10 Relay 1 normally open
9 No connection
8 No connection
7 Diaphragm zero adjust
6 Signal ground
®
Convectron
ATM Vacuum Gauge Module Instruction Manual
2