• Instruction manual WARRANTY The HI 731327 complete maintenance kit is also available, including: HI 93102 is guaranteed for two years against defects in workmanship • 2 glass cuvettes and materials when used for their intended purpose and maintained • Primary calibration standards: according to instructions.
• Battery status is monitored during every measurement cycle warning the user if the batteries become weak. In addition, HI 93102 will turn itself off before low voltage causes erroneous readings. • It utilizes a real time clock and recalls calibration data such as date, time and calibration values.
(light) of its TURBIDITY MODE molecules. HI 93102 has been designed to perform measurements according to Colorimetric analysis is based on the principle that specific compounds the USEPA’s 180.1 method and the Standard Method 2130B.
A substance absorbs a color complimentary to the color it emits. For FUNCTIONAL DESCRIPTION example, a substance appears yellow because it absorbs blue light. As a result, the Hanna meters use LED’s with specific wavelengths to measure samples. The optical distance (d) is measured by the internal diameter of the cuvette containing the sample.
Page 6
2) Secondary meter will quit diagnostic mode. If pressed LCD: The three-digit LCD shows the current for more then 3 seconds, an intermittent mode of measurement that is “F CL”, “CAL” prompt will appear on the “tCL”, “CY”, “PH”, “Id”, “Br”, “LFE”, upper LCD level and the calibration “tr”, and diagnostic or calibration modes, procedure is entered.
SPECIFICATIONS OPERATIONAL GUIDE y t i 0 . 0 e n i L / g SET CURRENT TIME/DATE L / g To set or change the current time, turn on the meter. After l a t L / g initialization routine, the LCD will show: L / g e n i 0 .
Page 8
Press READ. The meter will intermittently display “SIP” on the upper CURRENT TIME/DATE RECALL level of the LCD. To recall current TIME/DATE press and hold the ALT and READ keys together. The current time and an intermittent “TIME” will be displayed.
Page 9
The meter will show “SIP” for a few seconds and then a zero indication: Measuring Iron Fill one graduated mixing cylinder up to the 25 mL mark with deionized water. Add the content of one packet of HI 93746 TPTZ reagent, close the cylinder and agitate for 30 seconds.
Page 10
Add the content of one packet of HI 93746 TPTZ reagent, close the Note: Absorbance readings may have a negative sign if the cylinder and agitate for 30 seconds. transmittance of the “ZERO” sample is less then that of the reacted sample.
Add the appropriate reagent into the blank sample cuvette. Shake CALIBRATION and allow a few seconds for color to develop. TURBIDITY CALIBRATION The meter should be properly calibrated with a standard prepared as described in USEPA method 180.1. Span calibration To calibrate the span of the meter, fill the cuvette with the primary turbidity formazine standard of 50 NTU.
Page 12
If the CAL key is not pressed, the The display will blink “-Lt-” for upper display will show “----”, several seconds, indicating that the LED is being adjusted for the indicating that calibration mode turbidimetric channel. was not entered. In which case, hold down the ALT and CAL keys Afterwards, a sequence of numbers between -511 to 512 will appear together for 3 seconds to restart the...
Page 13
After the second calibration point is memorized, the unit will store the Using , , keys set the lower calibration data together with time and date in the EEPROM while part to “d 31”. intermittently indicating “CAL” and “Stor” for several seconds. Insert the previously prepared deionized water standard into the cuvette holder and make sure that the notch on the cap is positioned securely into the groove.
Page 14
After entering the calibration mode, Using the keys, set the the display will show the first point lower part of the LCD to “d 11” of the previous calibration with the and then press the key. most significant digit blinking. key is used as a toggle function in this mode and allows the Using the , keys, choose the concentration of the first...
“----”, indicating that the meter is calibrated and ready to measure the concentration of an HI 93102 facilitates operations by providing a diagnostic mode. In unknown sample. this mode, user can set or verify different parameters necessary to By pressing the CAL key during calibration, user can ensure optimum performance of the meter.
(memory). number, value, date and time. HI 93102 also reminds user if its memory is full. e.g. The first recorded reading in the buffer is lot # 0, 0.35 mg/L of Iron, memorized on 23 August at 3:34 pm;...
USER-SELECTABLE AUTO-OFF A “-BA-” indication will appear on the display when the batteries are too weak to perform accurate With HI 93102, the users can customize the shutdown time to save measurements. This message power. appears for a few seconds and...
Any variation introduced by the user to the supplied equipment may degrade the instruments' EMC performance. Hanna Instruments reserves the right to modify the design, To avoid damages or burns, do not put the instrument in microwave oven. For yours construction or appearance of its products without advance notice.