Specifications - Campbell 108 Product Manual

Temperature probe
Hide thumbs Also See for 108:
Table of Contents

Advertisement

Compatible CRBasic data loggers: CR6, CR3000, CR1000X, CR800-series, CR300-series,
l
CR1000

6. Specifications

Sensor element:
Survival range:
Measurement range:
Time constant in air:
Maximum cable length:
1
Accuracy
Worst case:
Interchangeability error:
Steinhart-Hart equation error:
Probe weight and dimensions
Weight with 10 ft cable:
Length:
Diameter:
Compliance:
1
Overall probe accuracy is a combination of thermistor interchangeability, bridge-resistor accuracy, and error of the
Steinhart-Hart equation. Bridge resistors have 0.1% tolerance with a 10 ppm temperature coefficient.
Interchangeability is the principle component error. If needed, an estimate of the interchangeability error for 0 to
50 °C, that can be used as the Offset parameter of the
or 2-point calibration.
Measurement Specialties 100K6A1iA thermistor
–50 to 100 °C
–5 to 95 °C
200 ± 10 s
305 m (1000 ft)
±0.3 °C (–3 to 90 °C)
±0.7 °C (–5 to 95 °C)
±0.10 °C (0 to 70 °C)
±0.14 °C at –5 °C
±0.25 °C at 85 °C
±0.35 °C at 95 °C
≤ ±0.01 °C (–35 to 50 °C)
136 g (5 oz
10.4 cm (4.1 in
0.762 cm (0.3 in
View the EU Declaration of Conformity at
ww.campbellsci.eu/108
Therm108()
instruction, can be determined with a 1-point
108 Temperature Probe     5