  • Page 1 USER MANUAL Si-V V 3 VANE THERMO-ANEMOMETER www.sauermanngroup.com...
  • Page 2: Instrument Description

    1 - Instrument description The Sauermann Si-VV3 is a vane Vane probe with 2 m cable thermo-anemometer measuring instrument. LCD screen In addition to the measurement of air On/Off & backlight button velocity and temperature, the airflow is also calculated.
  • Page 3: Safety And Environment

    The application operation is under the exclusive customer or user entity responsibility, who acknowledges using this system at his/her own risks. The customer or user entity explicitly exclude Sauermann, and every other company through which the it could have been sold of any kind of responsibility...
  • Page 4 2. This device must accept any interference received, including interference that may cause undesired operation. © 2020 Sauermann. All rights reserved. Sauermann & Si-HVACR Measurement MobileApp are the exclusive property of Sauermann. Non contractual document. The products functionalities and visual appearances can be modified without prior...
  • Page 5: Technical Specifications

    4 - Technical specifications Measuring Parameters Accuracy* Measuring range Resolution units ±(3% of value +0.1 m/s) from 0.4 to 3 m/s ±(1% of value +0.3 m/s) 0.01 m/s from 0.4 to 5 m/s from 3 to 30 m/s m/s, km/h, From 0.4 to 30 m/s 0.1 m/s from 5 to 30 m/s Air velocity...
  • Page 6 Vane and Hot wire anemometer Vane and Hot wire anemometer 6 - Dimensions 70 mm 70 mm 33 mm 33 mm (2.76’’) (2.76’’) (1.3’’) (1.3’’) 38 mm 38 mm (1.5’’) (1.5’’) 37.90 mm 37.90 mm (1.46’’) (1.46’’) 7 - Accessories Name Reference Softcase...
  • Page 7 8 - Operating instructions • Press simultaneously ON/OFF & HOLD Insert the batteries • Unscrew the batteries compartment keys to activate or deactivate the cover at the back of the device and wireless communicaiton. remove the cover. • When activated, the wireless •...
  • Page 8: Maintenance

    9 - Maintenance Change batteries • With the device turned off. • Unscrew the battery compartment at the back of the device and remove the cover. • Remove used batteries and insert new batteries (3 alkaline batteries AAA LR03 1.5 V) respecting the polarity. •...
  • Page 9 Sauermann Industrie Sauermann Italia SA ZA Bernard Moulinet Via Golini 61/10 24700 Montpon 40024 Castel S.Pietro Terme (BO) France T. (+39)-051-6951033 T. +33 (0)5 53 80 85 00 F. (+39)-051-942254 services@sauermanngroup.com info.italy@sauermanngroup.com Sauermann NA Sauermann Ibérica 140 Fell Court, Ste. 302 C/Albert Einstein 33.