Spectrum analyzer
Table of Contents

Advertisement

2. Display Adjustment
25. Press (MKR), [m),
M
A
R
MARKER-, REF LVL again.
Press
(SAVE), SAVE STATE
27. Set the sweep time to 10 ms.
28. Press
29. Adjust A2R209 SWEEP OFFSET to place the beginning of the trace at the leftmost
vertical graticule line.
30. Adjust A2R271 SWEEP GAIN to place the end of the trace at the tenth vertical graticule
line (one division from the right edge of the graticule).
31. Press 6-1 and press the @) key seven times.
Press
(SAVE),
SAVE
Press
[RECALL), RECALL STATE
state 1 is lined up with the top line of the graticule and the state 2 trace is lined up with
the eighth graticule down from the top (counting the top line). Repeat until the traces
align to within f0.2 divisions.
36. Adjust A2R209 and A2R271 until the start of sweep is aligned to the leftmost vertical
graticule line and the end of the sweep is aligned with the right most vertical graticule
line.
37. Press STATE 2 and STATE 3. The two traces should be aligned within 4~0.1 divisions.
Press
STATE 0
2-20 Adjustment Procedures
R 4 REF LYL. If the marker is not at the top graticule, press
K
E
STATE 1.
S
2.
TATE
STATE
and STATE 1 . The two traces should be aligned within fO.l divisions.