Getting Started
First, make sure you have a compatible controller (computer), and the proper test equipment.
The following paragraphs describe requirements for controllers and test equipment. Once the
proper equipment is identified, proceed to "Equipment Connections."
The HP 85623 Frequency Response Adjustment software is only for the HP 85623 spectrum
analyzer. Errors will occur if this software is used to adjust spectrum analyzers other than
the HP 85623. The HP 85629B Test and Adjustment Module (TAM) is recommended for
automated frequency reponse adjustments of the HP 85603, HP 85613, and HP 85633.
The HP 85643/85653 Adjustment/Diagnostic Software is required to adjust the frequency
response of the HP 85643 and 85653 spectrum analyzers.
Frequency Response Adjustment software requires using any combination of one of the
following HP controllers and the HP BASIC operating system:
Other HP 9000 Series 300 controllers are compatible with the Frequency Response Adjustment
software; however, graph printouts might not be full width. This is especially true with
medium- and high-resolution displays.
The Frequency Response Adjustment program requires at least 1 MB of free memory. The
computer can have either single or dual HP-IB ports. Refer to "Dual-Bus Operation" under
"Frequency Response Adjustment Software" for information on using the program with dual
HP-IB ports.
3-2 Frequency Response Adjustment Software
Controller
HP 9000 model 216
HP 9000 model 236
HP 9000 model 310
Operating System
HP BASIC 2.0 with extensions 2.1
HP BASIC 3.0 and required BIN files
HP BASIC 4.0 and required BIN files