HP 8562E Service Manual page 54

Spectrum analyzer
Table of Contents

Advertisement

1. High Voltage Power Supply Adjustment
Digital multimeter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 3456A
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 34118A
After disconnecting the ac power cord, allow capacitors in the high voltage
supply to discharge for at least 30 seconds before removing the protective
cover from the A6 power supply.
1. Turn the spectrum analyzer off by pressing LuNE]. Disconnect the power cord and
remove the spectrum analyzer cover. Fold down the A2 controller, A3 interface, A4 log
amplifier/Cal oscillator, and A5 IF assemblies. Remove the A6 power supply cover.
2. Position the spectrum analyzer as shown in Figure 2-l. Connect the positive DVM lead to
3. Set the HP 3456A controls as follows:
Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4. Reconnect the power cord to the spectrum analyzer and press m to the on position.
5. Record the voltage marked on the A6Al HV module.
6. Adjust A6R410 HV ADJ for a voltage equal to the voltage recorded in step 5.
7. Press ILINE) to turn the spectrum analyzer off and disconnect the power cord. Wait at least
30 seconds for the high voltage power supply capacitors to discharge.
8. Disconnect the DVM test leads from A6TP401 and A6TP405. Reinstall the power supply
cover.
2-14 Adjustment Procedures
Voltage marked on A6Al
DCVOLTS
HV Module =
Vdc