Promax AC-726 User Manual page 25

Cable and antenna analyser
Table of Contents

Advertisement

Is the electrical
length of open
and short cal kit
same?
yes
yes
The return loss of 50 ohms cal kit need to be > 42.
The VSWR of open/short cal kit need to be >100.
If the user does not focus on phase and impedance characteristics, cal kit from
other manufacturers can be directly used (usually short and open cal kit have
same electrical length).
NOTE:
The electric length is measured at 1GHz frequency, the unit is degree.
25
Is
the
exact
electrical length of
open and short cal
kit known?
unknown
know
know
Calibration setting menu
Figure 22.:
Return
Loss/VSWR/cable
Loss measurement
support
support
support
Phase/Impedenc-e
measurement
do not support
support
support
April 2016