GENERAL
The Type CA Unit contains two identical
amplifier channels that can be electronically
switched either by the oscilloscope sweep or at
a free-running rate of approximately 100 kc.
When amplifier switching is accomplished by
the oscilloscope sweep, the two signals to be
With
Instrument
Type
541/541A,
545/545A,
555,581,
585
551
531/531A,
533/533A,
535/535A
536
532
AA
CHARACTERISTICS
compared appear on alternate sweeps. Because
the sweeps are identical, and time-delay char
acteristics of the two amplifier channels are
closely controlled, time comparisons accurate
within 1 nsec can be made.
Either amplifier channel can be used separately
without electronic switching, making the Type
AMPLIFIER TRAN IENT RE PON E
Risetime
INPUT
INPUT
ELECTOR
ELECTOR
switch
switch in
in any
either DC
position
position
DC to 24 M c
15nsec
16 nsec
DC to 22 M c
DC to 15 M c
23nsec
DC to 10 M c
35nsec
70nsec
DC to 5 M c
Characteristics - Type CA
Passband
INPUT
ELECTOR
switch in
either AC
position
2 cps to 24 M c
.2 cps to 22 M c
with P6000 or
P6017 Probe
2 cps to 22 M c
.2 cps to 22 M c
with P6000 or
P6017 Probe
2 cps to 10 M c
.2 cps to 10 M c
with P6000 or
P6017 Probe.
2 cps to 10 M c
.2 cps to 5 M c
with P6000 or
P6017 Probe.
2 cps to 10 M c
.2 cps to 5 M c
with P6000 or
P6017 Probe.