the 5XT directly to A channel input connector.
When the adjustments to A channel have been
completed connect the cable from the Type
105 to the B channel input connector and proceed
with the
adjustments
ee the charts below for the adjustments to
each channel. Note:
position is reached the 5XT attenuator must
be removed.
A Channel Adjustments
/
Adj for Optimum
volts
cm
quare Corner
w
i
tch
.1
C3311C
C3312C
.2
C3313C
.5
C3314C
1
C3315C
2
C3316C
5
C3317C
10
20
C3318C
B Channel Adjustments
Adj for Optimum
VOLT /CM
quare Corner
witch
C4311C
.1
C4312C
.2
C4313C
.5
1
C4314C
C4315C
2
C4316C
5
C4317C
10
C4318C
20
14. H. F. Compensations
et oscilloscope controls as follows:
TIME/CM
et plug-in controls as follows:
VOLT /CM (A and B)
MODE
Connect the Type 107
ator to the channel B input connector of the
Type CA
Unit, installing the Terminating
Resistor between the P52 cable and the plug-in
unit. Adjust the Approximate Amplitude control
on the Type 107 for 3 cm of vertical deflection
on the crt of the oscilloscope. Adjust the
Approximate Frequency control for an approxi
mate output frequency of 450 kc.
Now adjust L3364, L4374, L4362, L4372,
L4334 and L4354 for an optimum square
comer with no overshoot on the square-wave
signal.
AA
for this
channel.
W h en the .5 VOLT /CM
Adj for Optimum
Flat Top
C3311B
C3312B
C3313B
C3314B
C3315B
C3316B
C3317B
C3318B
Adj for Optimum
Flat Top
C4311B
C4312B
C4313B
C4314B
C4315B
C4316B
C4317B
C4318B
.2 MICRO EC
.1
B ONLY
quare-Wave Gener
Calibration - Type CA
et the MODE switch of the Type CA Unit
to A ONLY and connect the Type 107, in the
same manner as above, to the A channel
vertical input connector. For channel A adjust
L3362, L3372, L3354 and L3334 for an optimum
square comer with no overshoot on the square
wave signal.
Note
When you are finished check to
see that the plugs in the coils are
all approximately the same depth
inside the coil form. If they aren't
readjust
until this condition is
obtained.
15. Frequency Response
The oscilloscope controls should be set as
follows .
TRIGGERING MODE
TIME/CM
TABILITY
The plug-in controls will be set to:
VOLT /CM (A and B)
MODE
Ad j ust the Type 190 Constant Amplitude
ignal Generator to obtain an output frequency
of 50 kc. Attach the Type 190 to the B channel
vertical input connector installing the 5XT
between the 190 attenuator box and the plug
in. Adjust the Output A mpl itude control on the
Type 190 to obtain exactly 3 cm of vertical
deflection on the crt. Adjust the Type 190
for an output frequency of 24 m c making sure
not to move the Output Amplitude control.
There should now be at least 2.1 cm of vertical
deflection remaining.
Turn the MODE switch to A ONLY and
connect the Type 190 to the B channel vertical
input connector. Make sure not to disturb the
control settings of the Type 190 when making
this transfer. The A channel should also have
at least 2.1 cm of vertical deflection.
AC LF REJECT
.1 MILLI EC
full right (cw)
.05
B ONLY
5-5
Need help?
Do you have a question about the TYPE CA and is the answer not in the manual?