Quick Start; First Operation - KEM TRICOR TCE80 -E Series Manual

Coriolis mass flow meter
Table of Contents

Advertisement

Getting Started

2.4 Quick start

WARNING!
As for safety and accuracy reasons many precautions must be taken, read chapter 3 carefully
before installing the unit!
In case the unit has only to be operated without flow for testing or learning purpose, the following
connections have to be made (see chapter 3.2):
Connect the supply voltage
Connect the TCE to the TCM (only required with the remote version)
The frequency and analogue outputs as well as the interface may be connected as well, if those features
are required
WARNING!
If the unit is connected to a bigger system, for your personal safety connect the protective ground
as well!
WARNING!
In hazardous areas it is not allowed to operate the unit without proper wiring according to chapter
3.3 and with the housing not properly closed!
2.4.1

First Operation

Make sure that all mechanical and electrical connections are made properly.
Switch on the power supply. The LED "OK" will flash green.
After the power up sequence the display shows the preselected values (ex factory normally flow and batch)
Switch on the flow. The value indicated in the display should be positive.
In case of an error the LED "ERR" will flash red.
As soon as the unit has reached the operating temperature, make the zero point calibration (see chapter
4.4):
Switch off the flow
Wait until the flow is zero
Start the zero point calibration in the SETUP menu
Wait until the offset procedure is finished
Switch on the flow again
The display can be altered by pressing the key "Display".
The error code, the device status and information like serial number, SW version a.s.o. can be viewed by
pressing the key "Info" for 3 seconds.
If the function is activated, the BATCH reading can be reset to zero by pressing the key "Reset".
For entering the CONTROL menu press "P" for 3 seconds.