oticon miniRITE R Instructions For Use Manual

oticon miniRITE R Instructions For Use Manual

Smartcharger

Advertisement

Quick Links

Instructions for use
Instructions for use
SmartCharger
SmartCharger
miniRITE R
miniRITE R
Oticon More™
Oticon More™
Oticon CROS PX
Oticon CROS PX

Advertisement

Table of Contents
loading

Summary of Contents for oticon miniRITE R

  • Page 1 Instructions for use Instructions for use SmartCharger SmartCharger miniRITE R miniRITE R Oticon More™ Oticon More™ Oticon CROS PX Oticon CROS PX...
  • Page 2 Introduction to this booklet Intended use This booklet guides you on how to use and maintain your new Oticon Intended use The charger is intended to recharge the hearing aids and SmartCharger miniRITE R. CROS transmitter that contain rechargeable batteries.
  • Page 3: Table Of Contents

    Table of contents About Warnings Package contents General warnings Getting started More info Charger Troubleshooting Temperature conditions Conditions of use Prepare charger Warranty Charge the hearing aids Compliance statements Daily use Charging your hearing aids Charging the charger Power bank function Power and charge using other sources Drying your hearing aids Caring for your charger...
  • Page 4: Package Contents

    The packaging box contains the following elements: Power plug (varies from country to country) Charger Instructions for use Instructions for use SmartCharger SmartCharger miniRITE R miniRITE R Oticon More™ Oticon More™ Oticon CROS PX Oticon CROS PX Cable This booklet About...
  • Page 5: Charger

    Charger Temperature conditions What it is and does Temperature Charging condition USB cable Power plug Below +5°C No charging due to low temperature. Insert in power plug For power socket (+41°F) +10°C to +35°C Best charging condition. The charging of hearing aids can (+50°F to +95°F) take up to 3 hours.
  • Page 6: Prepare Charger

    Prepare charger Connect to power source Charger is turned ON Step 1 Step 2 Step 3 Insert the USB plug Insert the other end of Insert the power plug When the charger is connected to into the power plug. the USB cable into the into a power socket.
  • Page 7: Charge The Hearing Aids

    Charge the hearing aids Place hearing aids in charger Charging status Place the hearing aids in the The hearing aid LED lights turn ON charging ports as shown in the when the hearing aids are correctly picture and close the lid. Make sure placed in the charger.
  • Page 8: Charging Your Hearing Aids

    Charging your hearing aids Ensure you fully charge your hearing aids before first use and charge them every night. That ensures that you start your day with fully charged hearing aids. If your hearing aid’s battery is completely drained, the normal charging time is: 3 hours 1 hour...
  • Page 9: Charging The Charger

    Charging the charger When the charger is connected to a power source, the LEDs on your charger indicate the charging status of the power bank. Your SmartCharger contains a built-in lithium-ion rechargeable battery that doubles as a power bank. This lets you fully charge a pair of hearing The LEDs turn OFF shortly after the power bank battery is fully charged.
  • Page 10: Power Bank Function

    Power bank function Power and charge using other sources When the charger is disconnected from power, the LEDs on your charger You can power and charge the charger’s battery from other sources than indicate the charger’s power bank battery level. You can check the a power socket.
  • Page 11: Drying Your Hearing Aids

    Drying your hearing aids Caring for your charger The SmartCharger uses heat from inductive charging to dry your hearing Always ensure that the charger is dry and clean before use. Keep the aids when you place them in the charger ports. charging ports clean from dust by using a dry cloth.
  • Page 12: General Warnings

    General warnings Choking hazards Explosives For your personal safety and to ensure battery. If your skin has been in contact correct usage, you should familiarize Charger, hearing aids and their parts The charger is safe to use under normal with the leaked battery acids, use a wet yourself fully with the following general should be kept out of reach of children and usage conditions.
  • Page 13 General warnings Fire hazards Transport on aircraft Active Implants Use of third-party accessories The power source in your charger has Consult the airline for details regarding The charger has been thoroughly tested Only use accessories, transducers, or insufficient energy to cause fire in normal transportation of your hearing aid charger and characterized for human health cables supplied by the manufacturer.
  • Page 14: Troubleshooting

    Troubleshooting Symptom Possible causes Solutions The hearing aids do not turn ON after being The charger did not have enough time to turn the hearing aids ON Place the hearing aids in the charger ports and try again, or use the push button to turn the removed from the before they were removed from the charger hearing aids ON manually...
  • Page 15 Troubleshooting Symptom Possible causes Solutions The hearing aid LED Depending on how depleted the battery in the hearing aid is, the hearing aid automatically blinks GREEN when the resumes normal charging mode after a timeframe of up to 10 minutes. The hearing aid has not been in use for an extended period of time hearing aid is placed in the charger...
  • Page 16: Conditions Of Use

    Conditions of use International warranty Your device is covered by an you have bought your device. Your Operating conditions Temperature: +5°C to +38°C (+41°F to +100°F) international warranty issued by the hearing care professional may also have Humidity: 5% to 93%, non-condensing manufacturer.
  • Page 17: Warranty

    Warranty Description of symbols and abbreviations used in this booklet Warnings Certificate Text marked with a warning symbol must be read before using the device. Manufacturer Name of owner: _______________________________________________________ The device is produced by the manufacturer whose name and address are stated next to the symbol.
  • Page 18 Description of additional symbols used on labels Humidity limitation Indicates the range of humidity to which the medical device can be safely exposed. Keep dry Indicates a medical device that needs to be protected from moisture. Firmware version used in the device. Caution symbol Consult instructions for use for warnings and cautions.
  • Page 19: Compliance Statements

    Compliance statements USA and Canada User information However, there is no guarantee that This device complies with part 18 This equipment has been tested and interference will not occur in a particular of the FCC Rules. found to comply with the limits pursuant installation.
  • Page 20 This medical device complies with Declaration of Conformity is available at the Medical Device Regulation (EU) the headquarters. 2017/745. Oticon A/S Kongebakken 9 DK-2765 Smørum Denmark www.oticon.global/doc SBO Hearing A/S Waste from electronic equipment must be handled according to Denmark local regulations.
  • Page 21 0000023229800001...

Table of Contents