Setting up the I/O card in the gas analyzer

Setting up the
I/O card in the
gas analyzer
Menu path
Serial number

Connecting the sample gas pipes

Sample gas
connections
Connect
sample gas pipes
22
The I/O card must be signed into the AO2000 gas analyzer. This takes place by
entering the serial number of the card in the AO2000 menu item Setup system
modules.
MENU  Configure  System  Setup system modules
The 14 digit serial number can be found on an adhesive label on the I/O card.
The serial number contains the following information (example):
0 0 6 0 1 6 0 0 0 1 2 3 0 1
The sample gas connections are PVDF screw connections for DN 4/6 mm hose.
The sample gas inlets and outlets on the top of the sample gas feed unit are
identified with arrows and numbers.
Connect the sample gas pipes to the sample gas inlets and outlets. The sample
gas pipes should be made from material that is suited to the measuring task.
Observe the sample gas inlet conditions (see page 8)!
SCC-F Sample gas feed unit Operator's manual
Serial number of the I/O card
Module type 006016 = I/O card
42/23-51 EN Rev. 3