Unscrew the safety screw at the upper
side end of the handle. Remove the
KaWe PICCOLIGHT F.O. otoscope
head from the handle and remove the
lamp. If necessary, clean the glass bulb
of the new lamp with alcohol. The glass bulb shall be clean and free
from fingerprints (grease spots). Put the lamp all the way in.
Unscrew the safety screw at the upper
side end of the handle. Unscrew the
KaWe PICCOLIGHT ophthalmo-
scope head from the handle. Remove
the lamp and insert the new incande-
scent lamp as far as it will go.
Technical data for KaWe MedCenter 5000:
Power supply unit with a plug ................. Type UES12LCP-050200SPA
Input voltage range ...................................... ..100–240 V~, 50/60 Hz
Output voltage ...........................................................................5.0 VDC
Output current .........................................................................max. 2 A
Cord length ...................................................................................... 3 m
Protection rating ...................................................................................II
Base unit (with end frame):
Base unit dimensions (WxHxD) ...........................206 x 146 x 100 mm
Spiral cord effective length ............2800 mm (700 mm spiral length)
Weight .................................................................................about 810 g
Add-on unit:
Add-on unit dimensions (WxHxD) ........................80 x 146 x 100 mm
Spiral cord effective length ............2800 mm (700 mm spiral length)
Weight .................................................................................about 370 g
Ear funnel unit:
Ear speculum unit dimensions (WxHxD) ...............80 x 146 x 100 mm
Weight .................................................................................about 200 g
Environmental conditions during use:
Temperature ............................................................ +10 °C to +40 °C
Relative humidity ............................................................10 % to 75 %
Atmospheric pressure .................................................700 to 1060 hPa
Lagerbedingungen und Transport:
Temperature .......................................................... -10 °C bis +60 °C
Relative humidity ...........................................................10 % to 95 %
Atmospheric pressure ................................................. 700 to 1060 hPa
Warnings:
• The device is maintenance free. The housing can only be opened
at the end frame to accommodate add-on units. Repairs may only
be carried out by qualified personnel.
• Only two KaWe MedCenter 5000 handles can be used at a time!
• The device has no connection components.
• In case of visible damages, remove the plug from the outlet im-
mediately. The power supply is a separating device and must be
disconnected from the mains if there is visible damage to the
device.
• The power supply unit must always be accessible.
• The wall station is not intended for operation in hazardous areas
and under the influence of intense magnetic fields, as well as in
operating rooms.
• Take care not to trip over the spiral cord. Use on escape routes is
prohibited.
• Only KaWe components may be used. It is forbidden to modify
the design of the device.
• During the examination, the user must not touch live parts such
as the plug or screw lock of the diagnostic device.
• If the lamps are lit for a long time (> 1 min), the metal surface of
the handle becomes very hot, which, if it comes into direct
contact with the skin, may cause burns, or if the handle is placed
on surfaces that are sensitive to high temperatures, it may catch
fire. Please note that after each use the handles must be returned
in their respective slots and the lights must be turned off. Allow
the handles to cool down after prolonged use.
• To prevent heating of the surface of the handles, the device auto-
matically turns off after 3 minutes. Activate it again by pressing a
button on the handle or by placing it in the wall console.
• When working with the wall station, special precautions regar-
ding electromagnetic copatibility must be observed.
• Portable and mobile high frequency telecommunication devices
may interfere with the operation of the wall station.
• The ear funnels of the ear funnel unit are for single use only and
must not be reused.
• The otoscope can be used to examine patients only with an in-
stalled ear funnel.
18