Error Messages
Error Message Descriptions
(16)
Input Overload Decrease max total power in
input.
Excessive input power has been detected which will
cause the ADC to clip the signal. Reduce the signal
level, change the attenuator/max total power setting
(under
Range
(17)
Data Acquisition forcing SHORT packing
The data acquisition rate is too high to use longer word
packing.
(18)
Command not implemented
The requested command is not implemented.
(19)
Function not implemented
The requested function is not implemented.
(20)
Signal exceeds maximum allowable power -
Reduce input power
Excessive input power has been detected which will
cause the ADC to clip the signal.
(21)
Memory Allocation FAILURE
(22)
Memory limit caused Data Acquisition to be
truncated
Caused by a Memory Allocation failure. The
measurement limited the acquisition time in order to
complete the measurement.
(23)
Setup INVALID
The parameters chosen create a measurement request
that is impossible to complete, often due to memory
limitations.
(24)
External reference missing
The external frequency reference signal either is
missing, has too low an amplitude, or does not match
the frequency value previously entered into the
instrument memory by the operator.
(25)
Even Second Clock missing
The even second clock signal supplied from the base
station is missing. Check the external trigger input
connection where the even second clock signal is fed
into the instrument.