Language Reference
SENSe Subsystem
Adjacent Channel Power—Automatic Measurement Points
[:SENSe]:ACP:OFFSet:LIST:POINts:AUTO OFF|ON|0|1,
OFF|ON|0|1, OFF|ON|0|1, OFF|ON|0|1, OFF|ON|0|1
[:SENSe]:ACP:OFFSet:LIST:POINts:AUTO?
Automatically selects the number of points for the optimum
measurement speed.
Factory Preset
and *RST:
Mode
Offset A
Basic
On
Remarks:
You must be in Basic mode to use this command. Use
INSTrument:SELect to set the mode.
History:
Revision A.03.00 or later
Adjacent Channel Power—Relative Attenuation
[:SENSe]:ACP:OFFSet:LIST:RATTenuation
<rel_powr>,<rel_powr>,<rel_powr>,<rel_powr>,<rel_powr>
[:SENSe]:ACP:OFFSet:LIST:RATTenuation?
Sets a relative amount of attenuation for the measurements made at
your offsets. The amount of attenuation is always specified relative to
the attenuation that is required to measure the carrier channel. Since
the offset channel power is lower than the carrier channel power, less
attenuation is required to measure the offset channel and you get wider
dynamic range for the measurement.
You can turn off (not use) specific offsets with the
SENS:ACP:OFFSet:LIST:STATe command.
Factory Preset
and *RST:
Mode
Offset A
Basic
0 dB
−40 to 0 dB, but this relative attenuation cannot exceed
Range:
the absolute attenuation range of 0 to 40 dB.
Default Unit:
dB
Remarks:
Remember that the attenuation that you specify is
always relative to the amount of attenuation used for
222
Offset B
Offset C
On
On
Offset B
Offset C
0 dB
0 dB
Offset D
Offset E
On
On
Offset D
Offset E
0 dB
0 dB
Chapter 5