Page 1
Procedure for Item Number: 4588254, 4588268 AMP-25, AMP-25-EUR Calibration Procedure AMP-25, AMP-25-EUR Calibration Procedure Document Change Record Rev.001 initial release 02/05/2015 Rev.002 Updated Verification Table from Verification test point sheet 0 3/11/2015 Rev. 002 Page 1 of 7...
Page 2
Procedure for Item Number: 4588254, 4588268 AMP-25, AMP-25-EUR Calibration Procedure Periodic calibration is required to conduct correct measurement. This instrument needs periodic calibration at intervals of one year. Calibrate the instrument according to the following procedures. Cautions Read instruction manual and understand the use of the instrument and its accessory(ies) correctly before proceeding the calibration activities.
Page 3
Procedure for Item Number: 4588254, 4588268 AMP-25, AMP-25-EUR Calibration Procedure 2.4. Move the jaw as below, then record the reading. Current coil 2.5. If the reading from step 2.3. and 2.4. have the difference > 0.1, adjust VR52 and repeat step 2.3. and 2.4. until both readings keep the difference ≤ 0.1 2.6.
Page 4
Procedure for Item Number: 4588254, 4588268 AMP-25, AMP-25-EUR Calibration Procedure Component position layout Rev. 002 Page 4 of 7...
Page 5
4.5. Repeat step 4.3. and 4.4. for the next calibration point. 4.6. If EUT beeps twice when pressing HOLD button, it indicates the CAL point fail to save. Repeat step 4.3. and 4.4. If EUT still fails to save CAL data point, contact Amprobe for serving.
Page 6
Procedure for Item Number: 4588254, 4588268 AMP-25, AMP-25-EUR Calibration Procedure 7. Exit Calibration mode 7.1. Turn OFF the EUT to exit calibration mode. 8. Verification Test 8.1. Verify the accuracy of the data points in Table 1 according to the accuracy specifications in users manual.