Summary of Contents for ELECTROMATIC Check-Line RH1

  • Page 1 Calibration Guide RH1 RH2 RH5 RH6...
  • Page 2 General Information On the basis of this calibration guide you are able to recalibrate the measuring instruments humimeter RH1, RH2, RH5 and RH6 by using humidity standards and the calibration device. If the accuracy check performed before the recalibration shows that the accuracy is still high enough, you do not need to recalibrate your device. Depending on your accuracy you have two options to recalibrate or adjust your RHx.
  • Page 3 Put in the first gasket ring in the upper part of the calibration device. Push in the Sensor-sword in the upper part as shown at the picture. Now put the second gasket ring into the upper part. Lay in the textile pad in the bottom part of the calibration device, and pour the humidity standard carefully at the textile pad.
  • Page 4 Conditioning the sensor To achieve best results, let the sensor condition for two hours. Read and annotate the measuring value Annotate the shown measuring value and also the temperature. Evaluation of the real humidity At ideal temperature of 23°C (measuring instrument, calibration device, calibration liquid) the reference value of the calibration liquid is valid.
  • Page 5 Enter the calibration Values into the MS-Excel-calculating sheet Enter all relevant values, these are the real humidity values and the old and new IC-values into the sheet. If there is a value missing, the calulation will be wrong. Enter the new calibration values into the RHx Do this as follow: 1.
  • Page 6 Trouble shooting If you recognize that there was something wrong with the calibration cycle, it is possible to reset the factory settings. ATTENTION: In this case all changes you have done will be reseted. Also your self calibrated moisture curves will be deleted.