Door Styles; Turn On The Product - Fluke CalPower 5128A Manual

Rhapid-cal humidity generator
Hide thumbs Also See for CalPower 5128A:
Table of Contents

Advertisement

Door Styles

(Standard) Square door with five ports, silicone gasket, center
tightening screw, and four fasteners (one at each corner)
To install and use this door, push the door into the chamber opening until it
stops. Then hand tighten the center screw to seal the round silicone gasket to
the chamber.The four corner screws (1/4 in x 20 threads) are used to
precisely locate the port locations and to further tighten the door against the
black outer seal on the outside of the chamber.
(Accessory) Round door with five ports, silicone gasket, and center
tightening screw (accessory)
To install and use this door, hand tighten the center screw to seal the round
silicone gasket to the chamber. Loosen the center screw to remove the door
from the chamber opening. To re-insert the door, align the rear acrylic piece,
middle silicone gasket, and front acrylic piece and then place the assembly
onto the chamber opening. Hand tighten the center screw to provide a snug,
airtight seal. The UUT probes, and any external reference probe(s), are
inserted into one of the five ports by removing the cap and placing the UUT
through the cap and appropriate-sized silicone grommet and port opening.
(Accessory) Clear square door with no ports, no silicone gasket, and
four fasteners (one at each corner)
To install and use this door, use the four corner screws (1/4 in x 20 threads)
to fasten the door in place and tighten the door against the black outer seal
on the outside of the chamber. Use this door to calibrate data loggers that are
placed into the chamber. Use this door in conjunction with a shelf inside the
chamber. See Data Logger Calibration.
To avoid measurement issues, do not open the chamber door when
the chamber temperature is below the ambient temperature. Water
vapor from the atmosphere will condense inside the chamber. If
condensation forms inside the chamber, shut the Product off and
remove the power cord. Thoroughly dry the chamber with a clean
paper towel before further use. See Remove Condensation.

Turn On the Product

To turn on the Product:
1. Push the "1" side of the power switch on the rear panel.
2. Monitor the startup screen for errors as the Product powers up. If an error
shows, see Troubleshooting and Error Code. Contact Fluke Calibration for
unresolved errors, see Contact Fluke Calibration.
3. Wait for to disappear from the main screen before the temperature
and humidity are set.
Note
RHapid-Cal Humidity Generator
Operation
19