Rhapid-cal humidity generator
Hide thumbs Also See for CalPower 5128A:
Table of Contents

Advertisement

1
3
Item
Name
Causes the air to circulate in a certain manner that gives better temperature
and humidity uniformity inside the insert. The volume inside the insert is the
working area where the Product guarantees its performance per the Product
Mixing Insert
specifications. See the end of this manual for a template that outlines the
working area. One of the three tabs on the outer part of the insert must be in
the 6 o'clock position as shown in the illustration ().This position ensures the
best temperature and humidity uniformity within the working area.
The chamber where the mixing insert is placed. The UUT can be placed
Outer
directly into the outer chamber without the mixing Insert, but the temperature
Chamber
and humidity will need to be monitored with an external reference probe.