Probe Calibrate Screen; Bit Humidity Gen Screen; Error Code Screen; Firmware Version Screen - Fluke CalPower 5128A Manual

Rhapid-cal humidity generator
Hide thumbs Also See for CalPower 5128A:
Table of Contents

Advertisement

5128A
Operators Manual

Probe Calibrate Screen

The probe calibrate screen shows the probe offsets for temperature and humidity
and is used in the Product-adjustment process. Use the probe calibrate screen to
load in a temperature offset and humidity offset to adjust the Product chamber
output. Probe offsets are entered as part of the routine calibration process.
An example of the calibration offset screen:

BIT Humidity Gen Screen

The BIT humidity gen screen contains parameters associated with the internal
humidity generator and is only used by trained service personnel. Push and hold
down EXIT to return to the main screen.

Error Code Screen

The error code screen is shown and errors that the Product detects are
presented. See Error Codes for a list of error codes. Push and hold EXIT to
return to the main screen.

Firmware Version Screen

Product firmware version details are shown. Push and hold EXIT to return to the
main screen.
16

°
°
