Procedure for Models with Firmware Version
Calibration Adjustment Counter
The Meter contains a calibration adjustment counter. The counter is incremented each
time a Calibration Adjustment Procedure is completed. The value in the counter can be
recorded and used to show that no adjustments have been made during a calibration
cycle.
Use the following steps to view the calibration counter on the UUT.
1. While holding down on the UUT, turn the rotary switch from OFF to
UUT should display . Release .
2. Press once to see the calibration counter. For example, .
3. Turn the rotary switch to OFF.
Calibration Adjustment Password
To start the Calibration Adjustment Procedure, the correct four-digit password must be
entered. The password can be changed or reset to the default as described in following
paragraphs. The default password is .
How to Change the Password
1. While holding down on the UUT, turn the rotary switch from OFF to
UUT displays . Release .
2. Press once to see the calibration counter.
3. Press again to start the password entry. The UUT displays ????.
The following keys represent the digit indicated below when entering or changing the
password:
K = 9
4.
Press the four keys to enter the old password. If changing the password for the first
time, enter
5. Press r to change the password:
•
The UUT displays //// if the old password is correct.
•
If the password is not correct, the UUT emits a double beep, displays ???? and
the password must be entered again. Repeat step 4.
6. Press the four keys representing the new password.
7. Press to store the new password.
M
R
= 1
= 2
G
r
= 5
= 6
M
R
(1),
(2),
≥
2.0
H
= 3
= 4
h
= 7
= 8
J = 0
H
(3), and
(4).