Procedure For Models With Firmware Version ≥2.0; Calibration Adjustment Counter; Calibration Adjustment Password; How To Change The Password - Fluke ProcessMeter 789 Calibration Manual

Hide thumbs Also See for ProcessMeter 789:
Table of Contents

Advertisement

Procedure for Models with Firmware Version

Calibration Adjustment Counter

The Meter contains a calibration adjustment counter. The counter is incremented each
time a Calibration Adjustment Procedure is completed. The value in the counter can be
recorded and used to show that no adjustments have been made during a calibration
cycle.
Use the following steps to view the calibration counter on the UUT.
1. While holding down  on the UUT, turn the rotary switch from OFF to
UUT should display . Release .
2. Press  once to see the calibration counter. For example, .
3. Turn the rotary switch to OFF.

Calibration Adjustment Password

To start the Calibration Adjustment Procedure, the correct four-digit password must be
entered. The password can be changed or reset to the default as described in following
paragraphs. The default password is .

How to Change the Password

1. While holding down  on the UUT, turn the rotary switch from OFF to
UUT displays . Release .
2. Press  once to see the calibration counter.
3. Press  again to start the password entry. The UUT displays ????.
The following keys represent the digit indicated below when entering or changing the
password:
K = 9
4.
Press the four keys to enter the old password. If changing the password for the first
time, enter
5. Press r to change the password:
The UUT displays //// if the old password is correct.
If the password is not correct, the UUT emits a double beep, displays ???? and
the password must be entered again. Repeat step 4.
6. Press the four keys representing the new password.
7. Press  to store the new password.
M
R
= 1
= 2
G
r
= 5
= 6
M
R
(1),
(2),
2.0
H
= 3
= 4
h
= 7
= 8
J = 0
H
(3), and
(4).
ProcessMeter™
Calibration Adjustment
V
. The
V
. The
33