Calibrator
Range
Resistance
400 Ω
120 Ω
400 Ω
300 Ω
4 kΩ
1.2 kΩ
4 kΩ
3 kΩ
40 kΩ
12 kΩ
40 kΩ
30 kΩ
400 kΩ
120 kΩ
400 kΩ
200 kΩ
400 kΩ
300 kΩ
4 MΩ
1.2 MΩ
4 MΩ
3.0 MΩ
40 MΩ
12 MΩ
40 MΩ
30 MΩ
Table 8. Resistance Measurement Test
Calibrator
Compensation Mode
2-Wire
2-Wire
2-Wire
2-Wire
2-Wire
2-Wire
OFF
OFF
OFF
OFF
OFF
OFF
OFF
Minimum
Maximum
Reading
Reading
119.6 Ω
120.4 Ω
299.2 Ω
300.8 Ω
1.197 kΩ
1.203 kΩ
2.993 kΩ
3.007 kΩ
11.97 kΩ
12.03 kΩ
29.93 kΩ
30.07 kΩ
119.7 kΩ
120.3 kΩ
199.5 kΩ
200.5 kΩ
299.3 kΩ
300.7 kΩ
1.193 MΩ
1.207 MΩ
2.986 MΩ
3.014 MΩ
11.67 MΩ
12.33 MΩ
29.22 MΩ
30.78 MΩ
ProcessMeter™