Automated Cleaning Instructions with Manual Precleaning
7
Load washing machine basket
Place articles in the specially designed trays for machine
washing supplied by Synthes. Ensure that all cannulations
(handpiece and attachments), if applicable, are positioned
vertically, i.e., in an upright position as shown.
Note: Never immerse the handpiece, lid or attachments in
an ultrasonic bath or any other liquid as this could decrease
the service life of the system.
8
Mechanical cleaning cycle parameters
Minimum
Cycle
time (minutes)
Rinse
2
Prewash
1
Cleaning
2
Rinse
5
Thermal disinfection
5
Dry
40
9
Inspect devices
Remove all devices from washing basket. Inspect the
cannulations, coupling sleeves, etc. for visible soil. If
necessary, repeat the manual preclean/mechanical cleaning
cycle. Please refer to Synthes Battery Powered Equipment
Lubrication Chart after cleaning and prior to sterilization.
© 2009 Synthes, Inc. or its affiliates. All rights reserved.
Temperature / water
Cold tap water
Warm tap water (40ºC)
Warm tap water (45ºC)
DI (Deionized) water
Hot DI water (93º C)
90ºC
Synthes (USA)
Synthes (Canada) Ltd.
1302 Wrights Lane East
2566 Meadowpine Boulevard
West Chester, PA 19380
Mississauga, Ontario L5N 6P9
Telephone: (610) 719-5000
Telephone: (905) 567-0440
To order: (800) 523-0322
To order: (800) 668-1119
Fax: (610) 251-9056
Fax: (905) 567-3185
Synthes is a trademark of Synthes, Inc. or its affiliates.
Type of detergent
N / A
Neutral pH enzymatic cleaner
Neutral pH enzymatic cleaner
N / A
N / A
N / A
www.synthes.com
Printed in U.S.A. 5/11 J9539-B

Advertisement

loading
Need help?

Need help?

Do you have a question about the Trauma Recon System and is the answer not in the manual?

Questions and answers

This manual is also suitable for:

Trs