Caution!
The zero point correction sets the current ACT value to
zero. In order exclude erroneous measurements, ensure
no system pressure is being applied when carrying out
this function.
This initiates the zero point correction. The ACT (actual)
value is indicated in the display as 0.0 bar. The
correction remains active until the device is turned off.
OFL/ZEro
is
seconds if the measured pressure
(0 bar) is greater than 5% of the
measurement range.
Zero point correction cannot be
carried out. Please ensure that no
system pressure is being applied.