Memory Mode Test (1653B and 1654B Only)

Use the following test to verify the proper function of the UUT Memory Mode:
1. Turn the rotary switch to V.
2. Connect the PE and L input jacks of the Tester to the 5320A V, HIΩ, mA~ Output
terminals as shown in Figure 4.
3. Set the 5320A for the Volts function and apply 25 V, 50 Hz to the Tester.
4. Press UUT
5. Use
and the to set the data identity.
6. Press
to save the data.
To recall the saved data:
7. Press
8. Use
and to set the data identity.
9. Press
twice to recall the data. The recalled data should match the saved value.
Figure 4. Touch Pad Sense Test
to enter Memory Mode.
to re-enter Memory Mode.
PE
(L2/Green)
L
(L1/Red)
Electrical Installation Tester