Sensing Head - DALEMANS DAT 420 Instruction Manual

Gas or oxygen detecto
Hide thumbs Also See for DAT 420:
Table of Contents

Advertisement

2.2.

Sensing head

The sensing element of the DAT 420 gas detector is an electrochemical sensor operating accord-
ing to the amperometry principle. The sensor detects gas by producing a chemical reaction be-
tween gas and oxygen contained in the sensor. This reaction produces a small current which is
proportional to the concentration of the gas. That small current is then measured and converted by
the 4-20 mA transmitter of the DAT 420.
The sensing head of the DAT 420 consists of:
A flameproof "d" sensor housing (base + body + top).
A sintered metal filter.
A sensor gasket.
An electrochemical sensor.
A 4-20 mA transmitter.
The sensor housing is a stainless steel assembly (base, body and top) which can be dismantled to
allow plug-in sensor or sintered metal filter replacement.
Electrochemical sensor
Wires
www.dalemans.com
Body
Base
Figure 3: sensing head
Top
4-20 mA transmitter
O-ring
Sintered metal filter
Sensor gasket
|