Calibration Of Detector; Calibration Mask, Gas Cylinder And Voltmeter; Figure 5 : Opening The Fph01 Sensing Head - DALEMANS DAT 420 Calibration Manual

Toxic gas and oxygen detector
Hide thumbs Also See for DAT 420:
Table of Contents

Advertisement

6.

CALIBRATION OF DETECTOR

To properly calibrate the detector, the following operations must always be performed in the correct order :
1. Adjustment of the detector zero setting (ZEROING);
2. Adjustment of the detector sensitivity (CALIBRATION).
Avoid touching the transmitter circuit or any of the electronic components during the entire adjustment
procedure on the 4..20 mA transmitter inside the sensing head.
BEFORE opening the detector junction box or its sensing head :
Ex
6.1

CALIBRATION MASK, GAS CYLINDER AND VOLTMETER

Calibration consists of creating a calibration atmosphere around the sensing head by injecting a reference
gas. To avoid contamination of the calibration gas by the ambient atmosphere or air draughts, always use
the calibration mask supplied by Dalemans (Figure 2 - Page 4) to inject nitrogen or calibration gas.
To apply nitrogen or calibration gas on to the detector:
1. Screw the pressure-regulator/flowmeter onto the gas cylinder and tighten the connection.
2. Undo the locking screw of the sensor housing using the 1.5 mm hexagonal key (OUT00000115).
3. Unscrew the body and top assembly of the sensor housing using the OUT00000113 spanner and
the 30 mm flat spanner (OUT00000114). The flat spanner is positioned between the base of the
sensor housing and the box.
4. Connect one end of the flexible hose to the flowmeter outlet and the other end of the flexible hose to
one of the end-fittings on the calibration mask.
1.5 MM HEX KEY
OUT00000115
DAT420_MAN02_EN.docx • V1R1
Check that the hazardous area has been declassified.
Check, using portable equipment, that there is no dangerous gas present in the
atmosphere of the working area.
Wipe the junction box and the sensing head with a DAMP cloth to minimize the risk of
electrostatic discharge.
LOCKING SCREW
OUT00000113

Figure 5 : Opening the FPH01 sensing head

DAT 420
www.dalemans.com
30 mm FLAT SPANNER
OUT00000114
SENSING HEAD
6

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents