CONTENTS Page INITIAL INSPECTION and ASSEMBLY MOUNTING PROCEDURE II. GENERAL INTRODUCTION III. USING THE JENCO MODEL 6309 POT FRONT PANEL (Key function illustration) MAIN DISPLAY MODE REAR CONNECTORS D. TURNING ON/OFF THE INSTRUMENT IV. MODEL 6309 POT MODES NORMAL MODE...
I. INITIAL INSPECTION and ASSEMBLY Carefully unpack the instrument and accessories. Inspect for damages made in shipment. If any damage is found, notify your Jenco representative immediately. All packing materials should be saved until satisfactory operation is confirmed. MOUNTING PROCEDURE 1.
III. USING THE JENCO MODEL 6309 POT A. FRONT PANEL 6309POT JENCO CAL. MODE VIEW CURSOR pH RECAL. 1. The [MODE/-] MODE or CURSOR key. 1a. In Normal mode this key will change the ORP display to ORP ABS(olute) or ORP REL(ative).
3. The [ ] UP key. During Calibration/Setting mode this key will increment the current blinking digit of the active parameter. 4. The [ /pH Recal.] DOWN or pH RECALIBRATE key. 4a.During Calibration/Setting mode this key will decrement the blinking digit of the active parameter.
Page 7
NORMAL or CHECK STATUS operation. IF no relay is ON then the “NORMAL” message will be displayed. IF any relay is ON then the “CHECK STATUS” will be displayed. CURRENT OUT - this will display the actual output of the 4-20 mA output.
triggered the respective relay settings. The rightmost character will indicate RELAY 1 if the display is pH or RELAY 3 if the display is ORP. The leftmost character will indicate RELAY2 if the display is pH or RELAY 4 If the display is ORP. If the display is ORP the annunciators will only indicate the unit selected at the RELAY/mA option in the ORP CALIBRATION .
Page 9
1. Connect the AC line to the rear of the instrument. The model 6309 POT can be used with 115 or 230VAC 50/60 Hz. Power consumption is 6 watts. Make sure the EARTH connector is connected to the earth lead of the AC power line.
2. MAKE SURE YOU CONNECT THE AC POWER CORD TO THE CORRECT AC TERMINALS. CONNECTING INCORRECTLY MAY DAMAGE THE UNIT PERMANENTLY. D. TURNING ON/OFF THE INSTRUMENT By just plugging the unit to a correct AC voltage the unit will be ready for use.
IV. MODEL 6309 POT MODES A. NORMAL MODE Turning ON the unit will always display main display mode. This instrument is designed to provide four distinct measurements: 1. Temperature - current temperature solution, which always displayed. 2. pH - the degree of acidity or alkalinity...
2. At MAIN DISPLAY pressing the [MODE/-] key will toggle between ABSolute ORP and RELative ORP. 3. You can change any blinking options or digit by pressing the ] or [ ] keys. 4. For options in digit format you need to press the [MODE/-] key to move the cursor to the next digit .
b. USER SETTING You will only see this page if the unit is not password locked. This page is just a WARNING, informing that all relays and analog output are frozen, and that you can calibrate and change the settings. NOTE: FROZEN MEANS ALL THE RELAYS AND THE ANALOG OUT WILL MAINTAIN THEIR LAST STATE UNTIL THE USER RETURNS TO MAIN DISPLAY MODE.
4. SLOPE - this is part 2 of the pH calibration process, this line will display the buffer to be used for SLOPE calibration, depending on the choice you made on BUFFER 2. (See Step by step pH Calibration.) EFFICIENCY -After saving the SLOPE a new efficiency will be displayed for about 4 seconds and then will move to the next page.
8. Clean the pH electrode and temperature probe with the de- ionized/distilled water. 9. Place the pH electrode and Temperature probe into buffer 2. ] key. A “WAIT” message will blink indicating that 10. Press the [ the instrument is waiting for a stable reading. The display will be locked to the buffer value corresponding to the temperature of buffer 2 (See IX.
the blinking digit; use the [MODE/-] key to select another digit and the [ ] key to save the new set point. 2.RELAY2 - The control action for this relay is changeable, you can choose “HI”-action or “LO” action. (In HI-action the relay will turn ON if the pH is greater than the RELAY2 set point, in LO-action the relay will turn ON if the pH is less than the RELAY2 set point, which is modified by the hysteresis value and hysteresis mode.)(See...
f. ORP CONTROL SETTING 1.RELAY3 - The control action for this relay is changeable, you can choose “HI”-action or “LO” action. (In HI- action the relay will turn ON if the ORP is greater than the RELAY3 set point, in LO-action the relay will turn ON if the ORP is less than the RELAY3 set point, which is modified by the hysteresis value and hysteresis...
g. CURRENT SETTING 1. TRANSMITTER OUTPUT – This option would let you choose if the current output type is pH LINEAR, pH ANTILOG or ORP LINEAR. If the user chooses “ORP LINEAR”, the ORP analog output will be based on the RELAY/mA UNIT.
ID/Address for unit. connecting multiple model 6309PDTF or other Jenco models for logging purposes then this ID/Address must be unique for each connected unit. This ID/Address is the same address that must be used by the PC program to communicate with this unit.
V. CONTROLLING THE RELAYS A. ISOLATION VOLTAGE The maximum isolation voltage of the relay output contacts is 1500 VDC. The voltage differential between the relay output contacts and the load should not exceed 1500 VDC. B. OUTPUT LOAD The current through the relay output contacts should not exceed 5 Amp at 115 VAC and 2.5 Amp at 230 VAC in order not to cause permanent damage to the relay contacts.
If the relay action is set to LOW and the hysteresis mode is EDGE, the relay will turn ON at [(RELAY SET POINT)], and will turn OFF at [RELAY SET POINT+ (hysteresis value)]. CAUTION : Setting any hysteresis value to zero may cause jitter and possibly damage the relay(s).
E. ORP RELAYS There are two independent relay channels for the ORP display. (see figure 1.). The hysteresis mode (center or edge, see figure 1.) and hysteresis value will be used by all ORP relays. The action of the ORP relays are dependent on set point, relay action (HIGH or LOW), hysteresis mode (Center or Edge) , hysteresis value and the current ORP display.
C. pH LINEAR TRANSMITTER (mA) OUTPUT The analog output will produce a linear analog output if the user selects this option (see figure 2). The analog output will be 20mA 20mA Adjustable Adjustable Bandwidth Bandwidth pH/ORP pH/ORP 4mA Set 4mA Set 20mA Set 20mA Set point...
D. ORP LINEAR (mA) OUTPUT The analog output will produce a linear analog output if the user selects this option (see figure 2). The analog output will be dependent on the ORP_4 mA setting, ORP_20 mA setting and the current ORP (ABS or REL) display. The ORP LINEAR output is based on the following equation: = 4mA+(16mA)*(D - ORP(4)) /(ORP(20) - ORP(4))
E. pH ANTILOG (mA) OUTPUT The analog output will produce an antilog analog output if the user selects this option (see figure 3). The analog output will be dependent on the pH _4 mA setting, pH _20 mA setting and the current pH display.
VII. RS485 INTERFACE OPERATION A. INTRODUCTION This instrument is using the standard RS485 MODBUS RTU protoc ol. Please read the “protocol.doc” in the accompanying disk to help you program an application that will communicate with the instrume nt. There is also a simple demo program that you can install to see the protocol in action.
VIII. ERROR DISPLAYS AND TROUBLESHOOTING Temperature DISPLAY Possible cause(s) display display unit [Action(s)] "OVER" "OVER" a. Temperature > 120.0°C. [Bring buffer/solution to a lower temperature.] [Replace temperature probe.] b. No temperature sensor. [Use a temperature probe.] "OVER" "UNDR" Temperature < -10.0°C. [Bring buffer/solution to a higher temperature.] "OVER"...
Page 29
Temperature DISPLAY Possible cause(s) display display unit [Action(s)] "UNDR" don’t care ORP (ABS or a. ORP ABS display < -2500 REL) [Bring solution to a higher ORP reading] don’t care “OVER” a. Temperature > 120.0°C. [Bring solution lower temperature.] [Replace temperature probe.] b.
IX. pH BUFFERS The temperature characteristics of pH calibration buffers 4.00, 4.01, 6.86,7.00, 9.18 and 10.01 are stored inside the instrument. The buffers used to calibrate the instrument must exhibit the same temperature characteristics as the stored values. TABLE 1. °C 4.00 6.86...
X. SPECIFICATIONS Range Resolution Accuracy -2.00 to 16.00 pH 0.01 pH ±0.01 pH ± 1 LSD Display Range Accuracy Resolution ORP Absolute mV -2500 to 2500 mV ±0.2 % of span 1 mV ORP OFFSET -3999 to 3999 mV ±0.2 % of span 1 mV Temperature Range Resolution...
Page 32
Controller Control type (five) ON/OFF control Relay output 5A at 115VAC or 2.5A at 220VAC Resistive load only GENERAL Keys Audio feedback in all keys Security protect 4-digit password Communication RS485 Power: 115VAC or 230VAC 50/60Hz Fuse 0.315Amp/250Vfast acting glass tube Ambient Temperature range 0.0 to 50.0 °C Display:...
Need help?
Do you have a question about the 6309 POT and is the answer not in the manual?
Questions and answers