JENCO 6312DTB/DTF Operation Manual

Do/temperature controller

Advertisement

Quick Links

Operation Manual
MODEL 6312DTB/DTF
DO/Temperature Controller
JENCO
ELECTRONICS, LTD.
MANUFACTURER OF PRECISION INSTRUMENTS

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 6312DTB/DTF and is the answer not in the manual?

Questions and answers

Summary of Contents for JENCO 6312DTB/DTF

  • Page 1 Operation Manual MODEL 6312DTB/DTF DO/Temperature Controller JENCO ELECTRONICS, LTD. MANUFACTURER OF PRECISION INSTRUMENTS...
  • Page 2: Table Of Contents

    CONTENTS Page INITIAL INSPECTION II. GENERAL INTRODUCTION III. USING THE JENCO MODEL 6312DTB/DTF A. FRONT PANEL B. NORMAL MODE DISPLAY C. REAR CONNECTORS D. TURNING ON/OFF THE INSTRUMENT IV. MODEL 6312DTB/DTF MODES A. MAIN DISPLAY MODE B. CALIBRATION/SETTING MODE C. HOW TO USE THE KEYS a.
  • Page 3: Initial Inspection

    I. INITIAL INSPECTION Carefully unpack the instrument and accessories. Inspect for damages made in shipment. If any damage is found, notify your Jenco representative immediately. All packing materials should be saved until satisfactory operation is confirmed. MOUNTING PROCEDURE 1. Make a cutout on any panel, with a thickness of 1/16 in. (1.5 mm) to 3/8 in.
  • Page 4: General Introduction

    The system also have two isolated 4-20mA analog output for DO & temperature each, offset and span configurable for the DO & temperature display. The model 6312DTB/DTF comes with a RS485 interface that can easily let the © user log all data with an IBM PC/AT compatible computer.
  • Page 5: Using The Jenco Model 6312Dtb/Dtf

    III. USING THE JENCO MODEL 6312DTB/DTF A. FRONT PANEL 6312DT 1. The [MODE /-] key. 1a. In Main Display mode this key will change the DO display to DO%, DO ppb or DO ppm. 1b. In Calibration/Setting mode this key will move to the next digit of the current active parameter.
  • Page 6: Normal Mode Display

    4. The [] DOWN key. 4a. During Calibration/Setting mode this key will decrement the blinking digit of the active parameter. 4b. During” DO CALIBRATION” the user can press this key to skip ZERO calibration. 5. The [] ENTER key. During Calibration/Setting mode, this key will save the current modified parameter and move to the next parameter on the page if the parameter is the last one in the page then it will move to first parameter on the next available page.
  • Page 7: Rear Connectors

    annunciator - this will be displayed if Calibration /Setting pages are password locked meaning the user can’t change the values unless the correct 4 digit number has been entered. 4. H1 (High action relay 1) or L1 (Low action relay 1) annunciator - this is the status of the DO Relay 1, if this is displayed then the Relay is ON.
  • Page 8: Turning On/Off The Instrument

    ON once the user plugs in the power cord to an AC power supply. 1. Connect the AC line to the rear of the instrument. The model 6312DTB/DTF can be used with 115 or 230VAC 50/60 Hz. Power consumption is 6 watts. Make sure the EARTH connector is connected to the earth lead of the AC power line.
  • Page 9: Model 6312Dtb/Dtf Modes

    IV. MODEL 6312DTB/DTF MODES A. MAIN DISPLAY MODE Turning ON the unit will always start in main display mode. This instrument is designed to provide four distinct measurements: 1. Temperature - current temperature of the solution, which is always displayed.
  • Page 10: How To Use The Keys

    C. HOW TO USE THE KEYS 1. At MAIN DISPLAY you need to press and hold the [CAL/VIEW] key for two seconds to change the display to VIEW SETTING PAGE or if the instrument is password locked the display will be PASSWORD CHECK page. 2.
  • Page 11: Password Check Page

    6. If you don’t need to change the current blinking option, just press the [] ENTER key to move to the next selection. 7. You also can press and hold [MODE/-] key for two seconds and it will move back the cursor to the previous option. a.
  • Page 12: Do Calibration Page

    4. STD. CAL or PROD. CAL: In this calibration you can choose first if you are doing a “STD. CAL” (Standard Calibration) or a “PROD. CAL” (Product Calibration). To accurately calibrate the model 6312DTB/DTF you will need the following information: i. The approximate pressure (in mBar) of the region in which you plan to take your dissolved oxygen measurements.
  • Page 13 (a). STD.CAL – The unit of this calibration will depend on the unit of DO at normal/main mode. Follow these steps to do a Standard Calibration: 1. If the DO probe is new and not yet connected to the meter, follow the DO probe manufacturer’s instruction on how to prepare the probe for first use and how to connect to the meter.
  • Page 14: Do Control Setting Page

    (b). PROD.CAL – You can only select PROD.CAL if you did not do a ZERO CAL first. If you are trying to do a PROD.CAL but started a ZERO CAL you need to go back to ZERO CAL and press the [] DOWN key to skip it. The unit of this calibration will depend on the unit of DO at normal/main mode.
  • Page 15: Do Ma Out Setting Page

    (See chapter V. CONTROLLING THE RELAYS ). User can use [] and [] keys to select the relay1 action and press [] key to save the setting. After you select the RELAY1 action you can now select the RELAY1 set point. Use [] and [] keys to change the blinking digit, use the [MODE/-] key to select another digit and the [] key to save the new set point value.
  • Page 16: G.temp. Control Setting Page

    4. RS 485 ID – this is the unique ID/Address for the unit. If you are connecting multiple model 6312DTB/DTF or other Jenco models for logging purposes then this ID/Address must be unique for each connected unit. This ID/Address is the same address that must be used by the PC program to communicate with this unit.
  • Page 17: Relay Output Load

    B. RELAY OUTPUT LOAD The current through the relay output contacts should not exceed 5 Amp at 115 VAC and 2.5 Amp at 230 VAC in order not to cause permanent damage to the relay contacts. This rating is specified for resistive loads only. C.
  • Page 18: Temperature Relay

    E. TEMPERATURE RELAY One relay channel is available for temperature display which has independent set point, action (see Figure. 1) setting (HIGH or LOW) and hysteresis value. The action of the Temperature relay is dependent on set point, relay action (HIGH or LOW), hysteresis value and the current Temperature display.
  • Page 19: Rs485 Interface Operation

    Note: 1. The range for 4mA and 20mA settings is 0.00 to 499.9%, 0.0000 to 59.999ppm, 0 to 59999ppb.. 2. The absolute difference of the 4mA and 20mA settings must be greater or equal to 1.0%, 0.100ppm, 100ppb or else the analog output will be disabled. 2 0 mA 2 0 mA 1 2 mA...
  • Page 20: Viii.error Displays And Troubleshooting

    turned on both the meter(s) and the computer, you are now ready to program a simple routine to read data from the instrument. Read the included file “Model 6312DTB(F) protocol.doc” to help you make a program to communicate with the Model 6312DTB(F). VIII.
  • Page 21: Specifications

    During power-on Unit has failed its RAM test. [Turn instrument OFF and back to ON again.] [Return for service. (see X.Warranty)] IX. SPECIFICATIONS Display Range Accuracy Resolution Dissolved O % air-sat 0 to 400.0% ±0.2 % of span 0.1% (meter only) Dissolved Depends ±0.2 % of span...
  • Page 22: Warranty

    If you purchased the item from our Jenco distributors and it is under warranty, please contact them to notify us of the situation. Jenco Service Department alone will determine if the product problem is due to deviations or customer misuse.
  • Page 23 Jenco Electronics, Ltd. 4F., NO. 80, Songde Rd., Xinyi Dist., Taipei City 110, Taiwan TEL: 886-2-2345-6188 FAX: 886-2-2345-6791 E-Mail: sales@jenco.com.tw Website: www.jenco.com.tw Shanghai Jenco Instruments, Ltd. 18 Wang Dong Zhong Road Sijing Town, Songjiang Shanghai, China TEL: 86-021-5761-9599 FAX: 86-021-5761-9598 E-Mail: sales@jenco.com.cn...
  • Page 24 机型 文件名称 使用说明书(英) 6312DTB/DTF 汪蕾 K6312DTB/DTF- 2013-04-15 文件编号 编制人 编制日期 版本号 核准人 核准日期...

Table of Contents