horiba LAQUAact D-73 Quick Manual

horiba LAQUAact D-73 Quick Manual

Portable ph meter
Hide thumbs Also See for LAQUAact D-73:

Advertisement

Quick Manual
Portable pH Meter D-73/D-74/D-75
CAL
MEAS
MODE
ENT
LIGHT
http://www.horiba-water.com
CODE : GZ0000336673A
January, 2017
© 2017 HORIBA Advanced Techno Co., Ltd.
DATA
SET

Preparation

① Prepare pH standard solutions
② Connect the electrode to the meter
Electrode connector
③ Press the power key of the meter
ENT
LIGHT
④ Remove the protective cap from the electrode,
and open the internal solution filler port
Remove the protective cap.
pH measurement
4
4
9
9
pH
pH
pH
pH
Temperature connector
Press
Open the internal solution filler port.
※Illustration is image of D-74

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the LAQUAact D-73 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for horiba LAQUAact D-73

  • Page 1 ④ Remove the protective cap from the electrode, and open the internal solution filler port http://www.horiba-water.com Open the internal solution filler port. Remove the protective cap. CODE : GZ0000336673A January, 2017 © 2017 HORIBA Advanced Techno Co., Ltd. ※Illustration is image of D-74...
  • Page 2 Calibration When the measurement sample is known to be acidic, perform two-point calibration with pH 7 and pH 4. Measurement When it is known to be alkaline, perform two-point calibration with pH 7 and pH 9. When the sample pH is not known, perform three-point calibration with pH 4, pH 7 and pH 9. ①...

This manual is also suitable for:

Laquaact d-74Laquaact d-75

Table of Contents