3. Trouble shooting

3.1 Trouble shooting table

Symptoms
No mist produced
3.1
No mist produced
3.2
No mist produced / fuse F2 checked
3.3
No mist produced / chamber or transducer
3.4
changed
No mist produced / power supply changed
3.5
Device works normally, no mist transported
3.6
to patient
Device works normally, no mist transported
3.7
to patient / fan checked
Device works normally, no mist transported
3.8
to patient / fan checked / check power on fan
Heated tubing won't get warm, Device is
3.9
working normally
Heated tubing won't get warm, Device is
3.10
working normally
LED or push-buttons failure / display changed
3.11
LED or push-buttons failure
3.12
18
LED indications
Action
Red
Green Yellow
Off
Off
Off
Check Power leads, Main switch and fuse on the
rear (T800L250)
Off
Off
Off
Open device and check fuse F2 (T250L250)
On
On
On
Change chamber or transducer,
then turn on device again
On
On
On
Open device check fuse F3 (T1L250)
if defective change board
On
On
On
Set and test dry-running protection
Off
On
On
Open device, remove debris,
check fan for smooth turning
Off
On
On
Open device, check power on fan if OK Change fan
Off
On
On
Change PC board
Off
On
On
check heated tubing and change if necessary
Off
On
On
Open device, check fuse F1
if OK check relay 3 (Rel.3)
Off
Off
Off
Change display board
Off
Off
Off
Change PC board