5.8 Settings for Service Functions
Function
Group
LOCK
CODE
PRES
REV'C
LST'A
SERV
Service
functions
SIM
SIM2
MAX'
MIN'
User Manual - ETS Series Digital Temperature Sensors
Function
Locking code
Change locking
code
Reset
Revision counter 0
Last device
status
Simulation output
1 or 2
Max. indicator
Min. indicator
Settings
Enter the locking code for enabling the
0
device.
Freely selectable code 1...9999.
0 = no locking
0
A locking code already assigned can only
be changed by first entering the old code
for enabling the device.
Resets all entries to the factory setting
NO
(not via XT-SOFT)
YES
Increases by 1 with each configuration
Displays the last device status to occur
0
≠ 0
OFF:
No simulation
OFF
OPEN:
OPEN
Switch output open
CLOS
CLOS:
3.5
Switch output closed
(if analog
3.5:
output
available)
Simulation values for analog output in
mA (3.5 / 4.0 / 8.0 / 12.0 / 16.0 / 20.0
/ 21.7)
Display of max. measured process value
0.0
Display of min. measured process value
0.0