A.13 TEMP
Measurement method
Channel
Probe type
Measurement site
Measurement range
Resolution
Accuracy
Updating cycle
Response mean time
Alarm response time
Unit
Alarm limit
T1/T2 high limit
T1/T2 low limit
TD high limit
A.14 IBP
Measurement method
Channel
Measurement range
Resolution
Accuracy
Update time
Pressure sensor
Sensitivity
Impedance range
Unit
Alarm limit
SYS ALM HI
MEAN ALM HI
DIA ALM HI
SYS ALM LO
MEAN ALM LO
DIA ALM LO
Thermistor method
Dual-channel
YSI-2.252 K
Body surface probe: armpit
Body cavity probe: oral, rectum
0~50 ℃
0.1 ℃
±0.1 ℃
About 1 s
<10 s
≤2 min
℃ or ℉
Range( ℃ )
(low limit +1)~50
0~(high limit-1)
0~50
Invasive and direct measurement
Dual-channel
-10~300 mmHg
1 mmHg
±2 % or 1 mmHg, whichever is greater.
About 1 s
5 uV/V/mmHg
300~3000 Ω
mmHg, kPa, cmH2O
Range(mmHg)
(low limit+1)~350
-50~(high limit-1)
129
Step( ℃ )
1
Step(mmHg)
1