Cannon CAV 4.1 Instruction & Operation Manual

Cannon CAV 4.1 Instruction & Operation Manual

Single-bath automatic viscometer
Hide thumbs Also See for CAV 4.1:
Table of Contents

Advertisement

Quick Links

CAV®4.1
CAV®4.1
CAV®4.1
CAV®4.1
Single-Bath
Automatic Viscometer
Instruction
& Operation
Manual

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the CAV 4.1 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Cannon CAV 4.1

  • Page 1 CAV®4.1 CAV®4.1 CAV®4.1 CAV®4.1 Single-Bath Automatic Viscometer Instruction & Operation Manual...
  • Page 3 CAV® 4.1 Operator’s Manual CANNON Instrument Company® 2139 High Tech Road State College, PA 16803...
  • Page 4 Copyright Copyright © 2021 CANNON Instrument Company®. All rights reserved. Trademarks CANNON® and the CANNON logo, CAV®, VISDISK® and VISCPRO® are registered trademarks of CANNON Instrument Company. Contact Address: CANNON Instrument Company 2139 High Tech Road State College PA 16803, USA Phone: 1-814-353-8000;...
  • Page 5: Table Of Contents

    Safety Precautions ........................3 Navigating the Manual ........................3 Getting Started ......................... 4 Unpack and Inspect ........................4 Setup the CAV 4.1 ......................... 5 Connect Hoses .......................... 6 Fill the Bath ..........................8 Drain the Bath ......................... 12 Connect Cables ........................15 Set Bath Temperature (Touchscreen) ..................
  • Page 6 VISCPRO II Software Interface ....................44 Manage User Accounts (VISCPRO II / CAV 4.1 GUI Touchscreen) ..........45 Log In ............................46 Change Password ........................46 Setup and Edit User Accounts ....................46 Logout ............................ 47 Connect VISCPRO II to the CAV 4.1 ..................... 47 VISCPRO II Instrument Set Up ..................
  • Page 7 Shipping Notification ......................67 List of Figures Figure 1 — Complete CAV 4.1 System Setup ..................5 Figure 2 — Tubing Connections (Rear Close Up) ................6 Figure 3 — Back of CAV 4.1 ....................... 8 Figure 4 — Filling and Draining the Bath ................... 9 Figure 5 —...
  • Page 8 Figure 52 — Connection Properties ....................37 Figure 53 — TCP/IP properties Figure 54 — Advanced Settings ........38 Figure 55 — CAV 4.1 Main Screen Menu ..................42 Figure 56 — Bath Screen ......................... 44 Figure 57 – VISCPRO Interface Window ..................44 List of Tables Table 1 —...
  • Page 9 Table 19 — Method Settings: Advanced ..................58 Table 20 — Method Settings: Drop Criteria ..................59 Table 21 — CAV Data Table Fields ....................60 CANNON Instrument Company® | Overview...
  • Page 10 | CAV 4.1 Operation Manual...
  • Page 11: Overview

    /s (cSt) and 10,000 mm /s (cSt) from 20 °C to 150 °C, with available options. The CAV 4.1 combines CANNON quality and reliability with modern design and unique features to enhance lab productivity, reduce costs, and improve data quality.
  • Page 12: Specifications / Compliances

    Specifications / Compliances Table 1 — Specifications and Compliances Specifications Details Model CAV 4.1 Benchtop Kinematic Viscometer Methodology ASTM D445/D446, ISO 3104/3105, IP 71 Applications Formulated oil analysis, hydraulic oil analysis, additive analysis, marine fuel testing, base stock analysis, light and...
  • Page 13: Safety Precautions

    Refer all service and repairs to qualified personnel. Navigating the Manual Since the CAV 4.1 uses both a touchscreen interface on the instrument and the VISCPRO II computer software application, the instructions are broken into sections for each type of controlling software. The header of each section notes CAV 4.1 GUI Touchscreen...
  • Page 14: Getting Started

    CAV® 4.1 Operator's Manual Getting Started This chapter covers all steps necessary to set up, calibrate, and run a sample on the CAV 4.1. The procedures are intended to be performed in the order written. These are basic, simplified procedures designed to familiarize an operator with the instrument, with more detailed instructions provided in following sections.
  • Page 15: Setup The Cav 4.1

    ¼ “ wrench ---- Setup the CAV 4.1 Initial setup of the CAV 4.1 involves connecting the hoses and cables, and filling the bath with silicone bath fluid. Figure 1 — Complete CAV 4.1 System Setup CANNON Instrument Company® | Getting Started...
  • Page 16: Connect Hoses

    6 mm blue FEP tubing (15 ft) Connect Hoses Assemble the CAV 4.1 on a secure surface with sufficient room behind the instrument to maintain a clear airflow and allow for unimpeded connection of the solvent inlet and waste outlet tubing.
  • Page 17 1. Place the CAV 4.1 on a solid, level workbench located near a MAINS AC power connection capable of providing sufficient current (10 A at 115 V ) to the CAV 4.1. Depending on the materials tested, solvent used, and laboratory regulations, the work area may also need to be equipped with a ventilation system.
  • Page 18: Fill The Bath

    Figure 3 — Back of CAV 4.1 10. Connect the waste tubing to the waste receiver. 11. Measure, cut, and install the 6 mm blue FEP tubing as exhaust tubing. Fill the Bath The temperature bath must be filled with the appropriate silicone fluid prior to operation. The viscosity of the fluid depends upon the temperature of the bath during most operations.
  • Page 19: Figure 4 - Filling And Draining The Bath

    Bath Drainage Receptacle Procedure 1. Turn off power to the CAV 4.1 instrument and let the bath cool to room temperature. 2. Lift the hood to access the top of the bath. 3. Remove the float from the top of the bath and set aside. This provides the opening to fill the bath.
  • Page 20: Figure 5 - Remove Float From Top Of Bath To Reveal Opening

    5. Using the funnel, fill bath with the appropriate silicone bath fluid to the appropriate fill line on the bath. The top line is for lower temperatures (100 °C) and the bottom line is for higher temperatures (135 °C and above). Refer to Figure 7. | CAV 4.1 Operation Manual...
  • Page 21: Figure 7 - Filling Bath With 10 Cst Silicone Fluid

    Figure 7 — Filling Bath with 10 cSt Silicone Fluid Top Fill Line Figure 8 — Bath Filled to Top Fill Line CANNON Instrument Company® | Getting Started...
  • Page 22: Drain The Bath

    2. Lift the CAV 4.1 hood to access the bath. It flips up, as shown in Figure 9. Figure 9 — Lift the CAV 4.1 Hood 3. Remove the front panel of the CAV 4.1 by gently pulling it away from the front of the instrument, as shown in Figure 10. Set the front panel aside.
  • Page 23: Figure 10 - Remove Front Panel And Set Aside

    Tube covers the end of the nozzle Figure 11 — Hose Inserted Properly onto Bath Drain (with Close Up) 5. Insert the other end of hose into a silicone fluid bottle or appropriate bath drainage container. CANNON Instrument Company® | Getting Started...
  • Page 24: Figure 12 - Hose Draining Into Fluid Container

    Figure 13 — Open Drain with Allen Wrench and Close-Up 7. When the bath drain is completely open, the drain plug will rise out of the bottom of bath and no longer be flush with the bottom of bath. | CAV 4.1 Operation Manual...
  • Page 25: Connect Cables

    2. Plug the power cord into the MAINS AC outlet at the work station. 3. (optional) Connect the CAV 4.1 to the network switch with one CAT cable. 4. (optional) Connect the network switch to the PC with a CAT cable and USB adapter.
  • Page 26: Calibrate Temperature (Touchscreen)

    Target temperature temperature temperature temperature Figure 16 — Main CAV 4.1 Screen 2. Click the Target Temperature field. A field entry screen opens as shown in Figure 18 and 19. Figure 17 — Temperature Settings Screen Figure 18 — Field Entry Screen 3.
  • Page 27: Figure 19 - Thermometer Access Hole

    4. From the main screen, click the current bath temperature button for the bath. This opens Temperature Calibration screen, as shown in Figure 21. Figure 20 — Temperature Calibration Screen 5. Click the Reference Temperature field. A field entry screen opens. CANNON Instrument Company® | Getting Started...
  • Page 28: Figure 21 -Green Checkmark For Entry Submission

    Figure 21 —Green Checkmark for Entry Submission 7. Select Calibrate, wait for the system to save the adjustment, and then click to return to the Back main screen. 8. Remove and clean the thermometer, and then store safely. | CAV 4.1 Operation Manual...
  • Page 29: Set Vacuum And Pressure (Components)

    Set Vacuum and Pressure (Components) Set the vacuum and pressure of the CAV 4.1 before running the instrument. Use the following procedure to access the vacuum and pressure knobs within the instrument. Table 8 — Tools and Materials Required Description Part Number Assembled CAV 4.1...
  • Page 30: Figure 23 - Access Pin On One Side

    Figure 24 — Pull Out Chassis while Pulling Up Both Access Pins 4. Pull latch on side panel to open the right bath control wing. When the latch is open, the bath control wing swings out. | CAV 4.1 Operation Manual...
  • Page 31: Figure 26 - Side Of Chassis With Latch, With Close Up Of Latch

    Figure 25 — Side of Chassis with Latch, with Close up of Latch 5. The bath control wing swings out, allowing access to the Vacuum and Pressure knobs. Adjust as necessary. Figure 26—Side View of Open Bath Control Wing CANNON Instrument Company® | Getting Started...
  • Page 32: Figure 28 - Vacuum / Pressure Gauge

    Vacuum Control Pressure Control Figure 28 — Front View of Bath Control Wing with Vacuum and Pressure Knobs 6. Repeat the steps in reverse order to slide the chassis back into place and commence operation. | CAV 4.1 Operation Manual...
  • Page 33: Set Vacuum And Pressure (Touchscreen)

    2. From the bath screen, click the Service button. This opens the Service menu options. Refer to Figure 31. Figure 30— Services Menu 3. Click Adjust Pressure/Vacuum. Follow the instructions prompted on the screen, as shown in Figure CANNON Instrument Company® | Getting Started...
  • Page 34: Adjust The Default Wash Settings (Touchscreen)

    Adjust the Default Wash Settings (Touchscreen) Verify and adjust the default wash settings prior to running a wash for the first time. The CAV 4.1 wash settings are defined in the method settings of the last Note: run test. When starting the CAV 4.1 from a complete shutdown, it does not reference the last run test and uses the default wash settings.
  • Page 35: Run A Wash (Touchscreen)

    Hence, prior to training the sensors or running tests, it is vital to first run a wash through the CAV 4.1. 1. Insert an empty vial into position 1 of the carousel.
  • Page 36: Run A Sample Test (Touchscreen)

    CAV 4.1 runs at that temperature again. Retraining for a particular temperature is only necessary when the viscometer tube is changed or if sensor performance changes over time. 1. Insert an empty vial into position 1 of the carousel.
  • Page 37: Review And Export Data (Touchscreen)

    Review and Export Data (Touchscreen) Once a test completes, the results can be exported to a USB drive in CSV format. 1. From the main screen, click the Results button. 2. Review the results. CANNON Instrument Company® | Getting Started...
  • Page 38 A field entry screen opens. 4. Enter the sample file name. Then click the green checkmark to submit it. Save As window opens. 5. Select the desired location to save the data file, then click Save. | CAV 4.1 Operation Manual...
  • Page 39: Sample Preheating

    The preheating option on the CAV 4.1 is controlled via tube settings and can be used on either carousel. Prior to starting the process of activating preheating, ensure that the carousel is filled with the provided metal sleeves.
  • Page 40 Sample Cup Heater Duty Cycle Figure 41 — Sample Cup Heater Duty Cycle 5. Press Save. If not saved, red lights appear on-screen to indicate changes have been made that are not saved. Refer to Figure 43. | CAV 4.1 Operation Manual...
  • Page 41 – do not touch. Do not allow any metal items (other than the metal vial sleeves) to touch the glass. 7. Go back to the main screen to exit the Preheating options. CANNON Instrument Company® | Sample Preheating...
  • Page 42: Sample Te Cooling

    If instrument is equipped with TE cooling option, follow the instructions outlined below in order to engage the CAV 4.1 in thermoelectric bath cooling. Always use proper laboratory procedures and caution when working with the CAV 4.1 instrument.
  • Page 43 Figure 47 — TE Cooler Indicator Lit In order to troubleshoot and verify the TE Cooler is working properly, open the hood to ensure the blue light is lit. If not, contact CANNON Instrument Company service for assistance at service@cannoninstrument.com or 814-353-800.
  • Page 44 Blue Indicator Light Figure 48 — Blue Indicator Light for Proper TE Cooling | CAV 4.1 Operation Manual...
  • Page 45: Connecting With Viscpro Ii

    VISCPRO and CANNON instruments. Figure 49 — Closed Network Settings Connect Equipment All CANNON CAV systems and the PC (with USB 2.0 to Ethernet adapter) connect to the network switch using standard CAT 5e/6 cables. Table 9 — Tools and Materials Required...
  • Page 46: Install Viscpro Ii

    3. Connect a CAT cable from the Ethernet adapter to the network switch. 4. Connect CAT cables from each CAV 4.1 to the network switch. Install VISCPRO II There are no configurable options in the VISCPRO installation routine. Insert the USB and follow the on- screen prompts.
  • Page 47 6. Right-click the connection and select Properties. The Connection Properties window opens. 7. Uncheck Internet Protocol Version 6 (TCP/IPv6). 8. Double-click Internet Protocol Version 4 (TCP/IPv4). The Internet Protocol (TCP/IP) Properties window opens. CANNON Instrument Company® | Connecting with VISCPRO II...
  • Page 48: Configure Firewall Rules

    Enter 1982 in the port field. • Select connection. Click Next. Allow the • Check all boxes: Domain, Private, and Public. Click Next. • In the Name field, enter Viscpro_IN. Click Finish. 4. Configure an outbound rule: | CAV 4.1 Operation Manual...
  • Page 49 Select Custom. Click Next. • Select All programs and click Next. • Change Protocol Type to ICMPv4. • Click Customize. • Select Specific ICMP types. • Check Echo Request and click OK. Click Next. CANNON Instrument Company® | Connecting with VISCPRO II...
  • Page 50: Configure Usb Power Settings

    CAV Registry Settings The following registry keys are set up at CANNON prior to shipping and do not need to be adjusted for the initial installation. However, when reconfiguring a network or adding an additional CAV 4.1 to your...
  • Page 51 Registry Key CAV 4.1 GUI Default Description value Address Instrument ID Instrument Instrument name Serial Number Serial Text string Instrument serial number Number EthernetPortNumber 1982 TCP port on which VISCPRO listens available CANNON Instrument Company® | Connecting with VISCPRO II...
  • Page 52: In-Depth Operation

    Common instrument settings, method templates, and equipment status are shared between the programs. The operating status of the CAV 4.1 is shown on the left-most button of the main screen menu of the CAV 4.1 GUI (touchscreen). This button toggles between...
  • Page 53 Displays current activity Preheater temperature Displays current preheater temperature Run / Stop Click to start or stop a run Actions Wash Click to run a wash Pause Click to pause a run or wash CANNON Instrument Company® | In-Depth Operation...
  • Page 54: Viscpro Ii Software Interface

    VISCPRO II is a Windows® software program comprising the general interface application and instrument/analysis-specific modules. A typical VISCPRO II interface displays an open instrument window, analysis log, application specific status window, and general information status windows, as shown in Figure 57. Figure 56 – VISCPRO Interface Window | CAV 4.1 Operation Manual...
  • Page 55: Manage User Accounts (Viscpro Ii / Cav 4.1 Gui Touchscreen)

    CAV 4.1 GUI Touchscreen) VISCPRO II and CAV 4.1 GUI user accounts are independent of each another. The VISCPRO II accounts are associated with the entire software installation and are not instrument specific. Both VISCPRO II and the CAV 4.1 GUI Touchscreen start with three default accounts, as described in Table 14.
  • Page 56: Log In

    Touchscreen: click Users > Change Manager Password. Because new accounts cannot be created using the CAV 4.1 GUI, this is the only option. 2. Type the new password into the appropriate fields. 3. Click or the select the green checkmark.
  • Page 57: Logout

    Users > Logout. Connect VISCPRO II to the CAV 4.1 VISCPRO II and the CAV 4.1 must be set up and communicating in order to attempt to establish a connection. Refer to Connecting with VISCPRO II. The preheat option is factory installed. This section is only applicable for Note: instruments with the preheating option installed.
  • Page 59: Viscpro Ii Instrument Set Up

    CAV 4.1 Touchscreen, touch the Local Mode/Remote Mode button to toggle the instrument state to Remote. The icon turns green to indicate that the CAV 4.1 is found and communicating with VISCPRO II. 3. In the Menu, click and then Instruments.
  • Page 60: Set Vacuum And Pressure (Viscpro Ii)

    Train the sensors whenever operating the CAV 4.1 bath at a new temperature. Once trained, the calibration values for a particular temperature are saved and will be recalled if the CAV 4.1 is set to run at that temperature again. Retraining for a particular temperature is only necessary when the viscometer tube is changed or if sensor performance changes over time.
  • Page 61: Run A Wash (Viscpro Ii)

    Run a Sample Test (VISCPRO This procedure details how to run and configure various kinematic viscosity tests on the CAV 4.1 using VISCPRO II software. 1. Verify that the CAV 4.1 is running and has achieved steady-state temperature in the bath.
  • Page 62 You can copy the complete set of entries from one tray to another. • Restore Tray from Archive—restores a previously saved tray. • Archive—opens the archive to allow deleting saved Delete Tray from trays. 6. Click Run. The window opens. Select Trays | CAV 4.1 Operation Manual...
  • Page 63: View Test Results (Viscpro Ii)

    7. Select the trays to run. Click The CAV 4.1 begins running the selected trays. Additional samples may be added to the trays until the run completes. • Pause—opens a selection window. Pausing a tray "Now" immediately halts all actions for that tray.
  • Page 64: Control Sample Cup Heaters

    40 °C to 50 °C 90 °C to 100 °C 50 °C to 60 °C 100 °C to 110 °C 60 °C to 70 °C 110 °C to 120 °C 70 °C to 80 °C > 120 °C 100% | CAV 4.1 Operation Manual...
  • Page 65: Advanced Setup

    Viscometer tube calibration values are associated with specific temperatures. A calibration must be performed any time the CAV 4.1 bath is operating at a new temperature. Once the tube is calibrated at a particular temperature, the instrument saves the C and E constant values. The values are then automatically recalled if the tube is run at that temperature again.
  • Page 66: Method / Test Settings

    Method / Test Settings Method templates are stored in the CAV 4.1 registry. When VISCPRO II software connects with a CAV 4.1 instrument, the method templates in the instrument are pushed to VISCPRO II. Any changes made to the method templates in VISCPRO II synchronize with the instrument.
  • Page 67: Table 18 - Method Settings: Wash

    Time (sec) that solvent A is injected after bulb empties Solvent B Injection Time Time (sec) that solvent B is injected after bulb empties Solvent C Injection Time Time (sec) that solvent C is injected after bulb empties CANNON Instrument Company® | Advanced Setup...
  • Page 68 Time (sec) for a bulb 2 sample between when the tube vent is opened to atmosphere and the sample efflux starts Bulb 3 Vent Drain Time Time (sec) for a bulb 2 sample between when the tube vent is opened to atmosphere and the sample efflux starts | CAV 4.1 Operation Manual...
  • Page 69: Adjust Default Wash Settings

    5. In the Date Filter tab, set the appropriate range for the data to be displayed. 6. In the Sample Filter tab, set the filters on the type of data to be displayed. 7. In the Report tab, select the data items to display (refer to Table 21): CANNON Instrument Company® | Advanced Setup...
  • Page 70 Adds the first 10 efflux times to the analysis {1,20} Efflux %d Adds the first 20 efflux times to the analysis Avg Efflux Average efflux time calculated using valid effluxes C constant of calibration E constant of calibration Saybolt Furol Seconds calculation | CAV 4.1 Operation Manual...
  • Page 71 8. Click OK. The software prompts user to save the analysis. • If selecting YES, the analysis is displayed after it is saved. • If selecting NO, the analysis displays, but is not saved. CANNON Instrument Company® | Advanced Setup...
  • Page 72: Maintenance And Service

    • UI updater: this software is actually the complete UI updater program (UpdateUI.exe) that runs on the CAV 4.1. It is used to install the UI software and firmware updates. It is updated by copying it over the existing UpdateUI.exe.
  • Page 73: Update The Ui Software

    8. After the CAV finishes its rebooting routine, power cycle the unit by completely turning it off using the power switch on the hood. Wait 10 seconds, and then turn it back on. Update the Controller Firmware Be careful not to touch any moving parts or live wires in the CAV 4.1. Warning: Note: A video of this procedure is available on CANNON’s YouTube channel.
  • Page 74: Update The Viscpro Ii Software

    Warning: ambient or a safe temperature. 1. Turn off the CAV 4.1 and allow the bath to cool to ambient/room temperature. 2. Drain the silicone bath fluid into a clean container. Save the fluid to refill the bath. 3. Raise the hood.
  • Page 75: Replace Power Supply

    19. Calibrate the new viscometer tube. Replace Power Supply The CAV 4.1 power supply is a switching regulated power supply designed to handle 90 – 240 V . To accommodate different locations, only the MAINS AC power cable needs to be changed.
  • Page 76: Warranty

    Company to conform to the weight, specifications and standards stated on the package. Returning a Product to CANNON Before returning a CANNON product for repair or service, make every attempt to identify the problem. If, after careful checking, the problem remains unidentified or unsolved, telephone CANNON Instrument Company (or the local service agent) to consult with a product specialist.
  • Page 77: Required Information

    Please contact CANNON before returning a product that could possibly contain hazardous material. Shipping Notification Products returned without CANNON’s prior authorization will not be accepted. The customer may be billed a testing fee if a product is returned to CANNON and found to be working properly. CANNON Instrument Company® | Warranty...
  • Page 79 CANNON INSTRUMENT COMPANY® 2139 High Tech Road | State College, PA 16803 | USA 800-676-6232 | 814-343-8000 | Fax: 814-353-8007 sales@cannoninstrument.com | www.cannoninstrument.com...

Table of Contents