7 Faults

7
Faults
26
EN
OXYWAY
If you are not able to clear an error message with the aid of the
table, you should contact the manufacturer
WEINMANN Emergency or your authorized dealer to have the
device repaired. To avoid serious damage, do not continue using
the device.
Fault
Leak at cylinder
connection
Leak at device
connection
Leak at safety valve
Bent indicator, needle
not correctly pointing to
0, bent connections
Cause
Damaged sealing ring
Damaged sealing ring
Excessive pressure rise in
pressure reducer
Mechanical damage
Remedy
Replace sealing ring
(see
"8.5 Replacing the
sealing ring", page
29).
Check sealing ring at
device connection and
replace if necessary.
Have the pressure
reducer repaired.