Weinmann Emergency MEDUMAT Easy Description And Instructions For Use page 35

Table of Contents

Advertisement

Audio response up to
serial number 7999
"Open oxygen cylinder"
"Check respiration and
select mode"
"Adjust settings"
"Connect patient"
"Demandflow mode"
"Mask ventilation mode"
"Tilt head back"
"Secure mask tightly"
"Tube ventilation mode"
"Ventilation pressure limit
45 mbar"
"Check airways and minute
volume"
"Device malfunction"
"Administer alternative
ventilation"
"Check pressure hose
system and gas supply"
"Rule out respiratory arrest
and check mask fit"
"Close oxygen cylinder"
"Check ventilation system
and settings"
"Selected language:
English" (Deutsch, Français,
...)
"Audio response is off"
Table 1:
Open oxygen cylinder valve
Depending on whether or not the patient is breathing, set
MEDUMAT Easy to one of the modes Demandflow (Page 27), mask
ventilation or intubation (Page 24).
Depending on patient weight, set respiration frequency and minute
volume (Page 23).
Connect patient to ventilator via ventilation hose and patient valve
using the patient mask or the connector of the tracheal tube.
Demandflow mode is selected.
Mask ventilation mode is selected.
While tilting the head back, use the E-C technique to seal the mask.
Tube ventilation mode is selected.
Maximum ventilation pressure for tube ventilation.
MEDUMAT Easy has measured excessive airway resistance. Check the
airways or adjust respiratory frequency and minute volume settings to
suit the patient (Page 23).
The device is faulty or the battery is failing. The device can no longer be
used for ventilation. You must therefore use another ventilation method
(Page 39).
MEDUMAT Easy has measured low pressure on the inlet side. Check
whether the O
cylinder still contains sufficient oxygen and that the
2
oxygen hose is not leaking, kinked or jammed.
MEDUMAT Easy can no longer detect a breathing pulse (trigger) in
Demandflow mode. Check the patient's breathing, and if necessary
switch to a different ventilation mode. Check the connections and
mask fit.
After switching off the ventilator, turn off the O
O
supply.
2
Disconnection: a pressure rise of 8 mbar is not achieved during the
inspiration phase under controlled ventilation. This is usually due to an
interruption of the ventilation system or to a low minute volume setting.
Check the connections or adjust the minute volume to suit the patient.
When selecting the language for the audio response, press the mask/
tube switch as many times as necessary until the desired language is
announced.
Confirmation that audio response is deactivated.
Meaning
.
slowly
cylinder or the external
2
Using the ventilator
35
EN

Advertisement

Table of Contents
loading

Related Products for Weinmann Emergency MEDUMAT Easy

This manual is also suitable for:

Wm 28000Wm 28100Wm 28050Wm 28020Wm 28010

Table of Contents