6.5 Ventilation with PEEP valve

6.6 Ventilation with filter

26
EN
Using the ventilator
the tidal volume set. Use suitable volume measuring
devices to check the tidal volume actually adminis-
tered.
Check respiratory parameters during ventilation.
If lung compliance diminishes during ventilation, the
ventilator will react by increasing the ventilation
pressure at constant volume.
Example of ventilation sequence before and after
lung compliance diminishes
A PEEP valve can be fitted to the expiration tube on
the patient valve with an adapter.
This valve makes it possible to use positive end-expi-
ratory pressure (PEEP).
Please see the PEEP valve instructions for details of
settings.
A conventional filter with standard 15/
22 connectors can be fitted on the inspiration tube
of the patient valve for hygiene purposes and to con-
dition the air for inspiration. This will increase both