Invacare AVIVA FX User Manual page 19

Power wheelchair base
Table of Contents

Advertisement

Power frequency (50/60 Hz)
magnetic field IEC 61000-4-
8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
The AVIVA FX Power Wheelchair is intended for use in the electromagnetic environment specified below. The customer or the user of the AVIVA
FX Power Wheelchair should assure that it is used in such an environment.
IMMUNITY test
IEC 60601 TEST
Conducted RF
3 Vrms
150 kHz to 80 MHz
IEC 61000-4-6
Radiated RF IEC
61000-4-3
80 MHz to 2,5 GHz
60101916-A
3 A/m
Guidance and manufacturer's declaration – electromagnetic immunity
Compliance level
LEVEL
3 Vrms
3 V/m
20 V/m
26 MHz to 2.5 GHz per
ISO 7176-21
30 A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
Electromagnetic environment – guidance
Portable and mobile RF communications equipment should be
used no closer to any part of the AVIVA FX Power Wheelchair,
including cables, than the recommended separation distance
calculated from the equation applicable to the frequency of the
transmitter.
Recommended separation distance
(3.5/ 3 V/m = 1.667)
26 MHz to 800 MHz (3.5/ 20 V/m = 0.175)
800 MHz to 2,5 GHz (7/ 20V/m = 0.35)
3 Electromagnetic Compatibility (EMC) Information
19