Download Print this page

Specifications - Microsensor MPM480 Operation Manual

Hide thumbs Also See for MPM480:

Advertisement

Products measurement tools type approval No. 15F139-61.

2 Specifications

Pressure Range:-100kPa...0kPa~10kPa...100MPa
Overpressure:1.5 times FS or 110MPa(Min. value is valid)
Pressure Type:Gauge, Absolute or Sealed
Accuracy: ± 0.25%FS (typ.)
Long-term Stability:≤±0.2%FS/year
Zero Temperature Drift: ±0.03%FS/℃ (≤100kPa)
FS Temperature Drift: ±0.03%FS/℃ (≤100kPa)
Operation Temperature: -30℃~80℃ -10℃~70℃(Cable)
Storage Temperature: -40℃~120℃ -20℃~85℃(Cable)
1
Power Supply
:15V~28VDC(intrinsic safe version supplied through safe
barrier)
Signal Output: 4mA~20mADC(2-wire current)
0mA~10/20mADC (3-wire current)
0/1V~5VDC (3-wire voltage)
Resistance:≤(U-15)/0.02Ω(2-wire current or 3-wire current)
≥10kΩ(Voltage output)
Pressure Port:M20×1.5 Male with Waterline Seal
or G 1/2" male with flush diaphragm (version P)
or Clamp (DN25) connection
Housing Protection: IP65
Electrical Connection: Special Plug or Cable 1.5m
Other related device: Intrinsically safe version shall be power supply via a safe
Note:○
For two-wire transmitter with LED or LCD watch head, the lower limit
1
±0.02%FS/℃(>100kPa)
±0.02%FS/℃ (>100kPa)
-10℃~60℃(Exia) -20℃~60℃(EXD)
barrier.
± 0.5%FS (max.)
3

Advertisement

loading
Need help?

Need help?

Do you have a question about the MPM480 and is the answer not in the manual?

Questions and answers