Quick Start Guide
7. Non-Standard Paint options may cause risk from electrostatic
discharge. Avoid installations that cause electrostatic build-up on
painted surfaces, and only clean the painted surfaces with a damp
cloth. If paint is ordered through a special option code, contact the
manufacturer for more information.
5.8
Brazil
5.8.1
E2 INMETRO Flameproof and Dust
Certificate: UL-BR 13.0535X
Standards: ABNT NBR IEC 60079-0:2013, ABNT NBR IEC 60079-1:2016,
Markings:
Special Conditions for Safe Use (X):
1. See product description for ambient temperature limits and process
temperature limits.
2. The non-metallic label may store an electrostatic charge and become
a source of ignition in Group III environments.
3. Guard the LCD display cover against impact energies greater than 4
joules.
4. Consult the manufacturer if dimensional information on the
flameproof joints is necessary.
5.8.2
I2 INMETRO Intrinsic Safety
Certificate: [Fieldbus]: UL-BR 15.0264X [HART
Standards: ABNT NBR IEC 60079-0:2013, ABNT NBR IEC 60079-11:2013
Markings:
See
Table 5-5
Special Conditions for Safe Use (X):
1. The apparatus must be installed in an enclosure which affords it a
degree of protection of at least IP20.
2. Non-metallic enclosures must have a surface resistance of less than 1
GΩ; light alloy or zirconium enclosures must be protected from
impact and friction when installed in a zone 0 environment.
32
ABNT NBR IEC 60079-31:2014
Ex db IIC T6...T1 Gb; T6...T1: (–50 °C ≤ T
50 °C ≤ T
≤ +60 °C)
a
Ex tb IIIC T130 °C; IP66; (–40 °C ≤ T
[Fieldbus]: Ex ia IIC T* Ga (–60 °C ≤ T
IIC T* Ga (–60 °C ≤ T
for Entity Parameters and Temperature Classifications.
≤ +70 °C)
a
®
]: UL-BR 14.0670X
≤ +** °C) [HART]: Ex ia
a
≤ +** °C)
a
April 2021
≤ +40 °C), T5...T1: (–
a