SAFETY The meanings of WARNINGS, CAUTIONS, AND NOTES are: WARNING A WARNING indicates when failure to follow the instructions could result in injury to personnel. CAUTION A CAUTION indicates when failure to follow the instructions could result in damage to equipment.
Page 4
At high concentrations, it can cause discomfort, or at sufficiently high levels, be dangerous. The PC-3 series of lamps produce minimal amounts of ozone during start-up and no ozone during operation. The radiation beneath the wavelength of about 240 nm, given off by the arc, is absorbed by the outer, colder gas layers and therefore, produce no ozone during operation.
Page 5
SAFETY WARNING EYE PROTECTION Always wear eye protection when working with or near UV equipment. Use goggles, safety spectacles (glasses) or a face shield to protect your eyes. • Goggles should completely surround and protect your eyes. Many goggles will also fit over regular glasses.
® The Light-Welder PC-3 is a special purpose UV curing lamp used for pinpoint curing of UV adhesives, coatings, and potting materials. It emits a 5 mm diameter spot of UV light from a liquid light guide. This guide can be hand-held for complete mobility or clamped into position on assembly equipment or work stations for repetitive operations.
** Measured with an EIT Spotcure Radiometer. Values in W/cm CAUTION With the filter installed, the PC-3 lamp emits only UV-A and UV-B light and a low amount of UV- C light. Without the filter, UV-A, UV-B, and UV-C, and visible light are emitted. Never look directly at the light source while the unit is in use.
INSTALLATION 1. Install lamp per LAMP REPLACEMENT/INSTALLATION (Page 9). 2. Connect power cord to rear of unit and plug into a grounded wall outlet. 3. Connect light guide to bezel mount by pressing down on lock and inserting guide. 4. Press power switch. Button will illuminate when power is on. 5.
LAMP REPLACEMENT / INSTALLATION WARNING BEFORE REPLACING THE LAMP, THE UNIT MUST BE “OFF” FOR A MINIMUM OF 5 MINUTES. THIS WILL ALLOW DANGEROUS CHARGES, WHICH ARE PRESENT FOR PROPER LAMP OPERATION, TO BLEED OFF. LAMP REPLACEMENT PROCEDURE 1. Allow the unit to cool before removing the lamp/reflector assembly. 2.
MAINTENANCE LIGHT GUIDE Operating temperature should not exceed 55°C. Ends of the guide should be kept clean to transmit as much light as possible. Cured adhesive can be removed with razor blade. Avoid sharp bends with the light guide since this reduces light output and damages guide. TM-021 PN 35933 Rev 05/06/97...
TROUBLESHOOTING CAUTION ONLY QUALIFIED MAINTENANCE PERSONNEL SHOULD ATTEMPT THE FOLLOWING PROCEDURES: Problem: Lamp Will Not Ignite Possible Cause Testing Corrective Action Improperly Fastened Visually inspect all input/output Secure all connections. Connections connections (i.e. power cord, lamp). Lamp Beyond Useful Life Replace with a known good Replace lamp/reflector lamp/reflector assembly and re-...
Page 13
Transmission loss In Light Compare light guide output against Replace light guide. Guide Too Great new light guide (or use the Dymax Light Guide Simulator - P/N 36987) to determine transmission loss. Transmission loss In Filter Compare filter output against new Replace filter.
Page 15
TROUBLESHOOTING (Figure 3) Open Circuit Voltage From Lamp Output Test Configuration: Remove Power Remove Lamp/Reflector Assembly From Unit Measurement Taken Across Lamp Output From Unit Oscilloscope Setting: DC Coupling 20 ms / Div 500V / Div Probe Connected Across Lamp Output (*Use a High Voltage Probe*) ...
Nothing in this bulletin is to be interpreted as a representation of freedom from domination of patents owned by others, or a license under a Dymax Corporation patent.