Change 1
Change the Measurement Uncertainty value of Test Signal Frequency Accuracy Test (Page 9-10) to the
following.
Test Signal
Frequency
1 k
Change 2
Change the Measurement Uncertainty value of Resistance Measurement Accuracy Test (Page 9-10) to the
following.
Standard
[ohm]
1 m
10 m
100 m
1
10
100
1 k
10 k
Copyright 2008 Agilent Technologies
C
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Agilent 4338B Milliohmmeter Operation Manual
Manual Change
Test Limits
[Hz]
[Hz]
±1
Test Signal
Test Limits
Current
[A]
10 m
±0.024 m
10 m
±0.054 m
1 m
±0.45 m
10 m
±0.0043
10 μ
±0.044
1 μ
±0.49
10 μ
±9.3
1 μ
±540
Test Result
Measurement
Uncertainty
± 0.000018 k
Test
Result
[ohm]
[ohm]
Agilent Part No. N/A
July 2008
[Hz]
Measurement
Uncertainty
[ohm]
±0.0081 m
±0.025 m
±0.18 m
±0.00020
±0.017
±0.15
±0.0019 k
±0.025 k