Change 1
Add TAR in Test Signal Frequency Accuracy Test (Page 9-10) to the following.
Test Signal
Frequency
[Hz]
1 k
Change 2
Add TAR in Resistance Measurement Accuracy Test (Page 9-10) to the following.
Standard
Test Signal
Current
[ohm]
[A]
1 m
10 m
10 m
10 m
100 m
1 m
1
10 m
10
10 μ
100
1 μ
1 k
10 μ
10 k
1 μ
Copyright 2008 Agilent Technologies
C
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Agilent 4338B Milliohm meter Operation Manual
Manual Change
Test Limits
Test Result
[Hz]
±1
Test Limits
[ohm]
±0.024 m
±0.054 m
±0.45 m
±0.0043
±0.044
±0.49
±9.3
±540
Measurement
Uncertainty
[Hz]
± 0.000018 k
Test
Measurement
Result
Uncertainty
[ohm]
[ohm]
±0.0081 m
±0.025 m
±0.18 m
±0.0020
±0.017
±0.15
±0.0019 k
±0.025 k
Agilent Part No. N/A
Jun 2009
TAR
57
TAR
2.9
2.2
2.6
2.1
2.6
3.3
5.0
21.9