The memory location indicated does not contain any data.
The test cannot proceed because desired pressures within the ear
have not been achieved. This occurs when:
the tip is not completely sealed in the ear canal.
l
the tip becomes dislodged during a test.
l
the ear canal volume is abnormally large
l
above 2.5~~ the OPEN
To alleviate this condition:
reposition the tip.
l
use a different tip.
l
increase pressure against the ear.
l
NOTE: If a leak condition occurs after -100 daPa pressure is
reached, the pattern will be displayed with a Save the Data Symbol.
BLANK MEMORY
LEAK
symbol will appear).
12