Electromagnetic immunity
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
IEC 60601 Test
Compliance
Level
Level
±8 kV contact
±8 kV contact
±2 kV, ±4 kV,
±2 kV, ±4 kV,
±8 kV ±15 kV air
±8 kV ±15 kV air
±2 kV for power
±2 kV for power
supply lines
supply lines
±0.5 kV, ±1 kV
±0.5 kV, ±1 kV
line(s) to line(s)
line(s) to line(s)
Electromagnetic Environment -
Guidance
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30 %.
ESD may cause the Docking
Station to stop functioning, upon
which it will automatically reboot,
indicated by the LED switching off
momentarily (5-10 seconds) .