Warranty The GENTEC TPM-300-CE Laser Power Meter carries a one year warranty (from date of shipment) against material and/or workmanship defects, when used under normal operating conditions. The warranty does not cover damages related to battery leakage or misuse. GENTEC will repair or replace, optionally, any TPM-300-CE which proves to be defective during the warranty period, except in the case of product misuse.
SAFETY INFORMATION Do not use the TPM-300-CE if the device or the detector look damaged, or if you suspect that the TPM- 300-CE is not operating properly. Appropriate installation must be done for water cooled and air-force cooled detectors. Refer to the instruction manual for more informations.
SAFETY INFORMATION ..........................III SYMBOLS ..............................III TABLE OF CONTENTS ..........................IV LIST OF ILLUSTRATIONS ..........................V INTRODUCTION ............................ 1 THE TPM-300-CE DIGITAL POWER METER..................2 2.1 Introduction............................. 2 2.2 Specifications ..........................2 2.3 Front Panel Description........................3 3.1 Probe Input Jack Information ......................7 3.2 Power Measurement Procedure ....................
INTRODUCTION The TPM-300-CE is the highest quality Laser Power Meter available on the market. In order to obtain the maximum performance from the TPM-300-CE, it is preferable to read the following manual carefully. The TPM-300-CE has been specially designed for laser maintenance applications and laser power spot checks, but its characteristics make it useful for general laser beam power and measurement applications.
THE TPM-300-CE DIGITAL POWER METER Introduction The TPM-300-CE is a digital/analog power meter designed to be used with any of the new PS Series detector heads (see section 3.1 for further details). No calibration is required when connecting the detector heads. The TPM-300-CE is already internally set to accept all the PS Series detector heads.
The mW indicator signals a digital meter readout in milliwatts. W light indicator: The W indicator signals a digital meter readout in Watts. TPM-300-CE Instruction Manual Revision 5 November 2008 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 10
Fig. 2-1 TPM-300-CE Front Panel TPM-300-CE Instruction Manual Revision 5 November 2008 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 11
Back Panel Description Refer to Fig.2-2 ON/OFF switch: The TPM-300-CE is switched ON or OFF by pressing the ON*OFF switch. When the TPM- 300-CE is not in use, switch it off in order to prevent battery leakage and to extend meter autonomy.
Page 12
Fig. 2-2 TPM-300-CE Back Panel TPM-300-CE Instruction Manual Revision 5 November 2008 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
TPM-300-CE monitor. The Gentec-eo inc V1 to V2 (100-16651) adapter can be used to permit connection of the PS-310 Series and PS-330 Series Version 1 power detector heads. Please contact your local GENTEC distributor or the nearest Gentec-eo inc office for further information (see WARRANTY page).
Power Measurement Procedure Install the power detector head on its optical stand. Connect the power detector head to the TPM-300-CE using the PROBE INPUT jack. Slide the latch to the left to lock the connector into place. Remove the power detector cap.
The parity can be set only from a computer or a terminal. Select one of the following: None (8 data bits), Even (7 data bits), or Odd (7 data bits). When you set the parity, you indirectly set the number of data bits. TPM-300-CE Instruction Manual Revision 5 November 2008...
Displays the current scale X mW or XW SENSITIVITY? Gives the detector sensibility X.XXXXXX V RESPONSETIME? Gives the detector responsetime X.XXX S TPM-300-CE Instruction Manual Revision 5 November 2008 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 17
Invalid command: Input command does not correspond with the command listing. Invalid scale: The selected scale does not correspond to the detector's input power range. TPM-300-CE Instruction Manual Revision 5 November 2008 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
MAINTENANCE Battery Charging As mentioned previously, the TPM-300-CE power meter is operated using 5 standard rechargeable nickel cadmium batteries. When the LOW BATT. indicator is lit, recharge the batteries by connecting the external power supply for 12 hours. TPM-300CE can be On or Off during this procedure.
Page 22
Directive(s) and Standard(s). Place : Québec (Québec) (Signature) 24 November 2000 Date : Michel Giroux (Full Name) President TPM-300-CE Instruction Manual Revision 5 November 2008 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...