too close to the exllaust filter vvill restrJct the air
and limit the
intake velocity . .
t.,
mi_ni.mum of 12"
of clearance on all sides and top of the cabinet is
necessary to achieve ideal conditions for testing.
If a free space of 12" above the exhaust filter is not
available, inaccuracies in exhaust flow
measurements mav result. Consult with our
Customer Service Department for the implications
of this, and for alternatives.
Installing the Cabinet
Installation of this cabinet should be carried out
in accordance with appropriate OSHA regulations
and those of other regulatory agencies having
jurisdiction.
1.
First, move the cabinet from the unloading
area to its intended location on dollies. The
overall dimensions in the shipping position
for the ·!-foot unit are: 34
:I
12"
deep x 53
3
I
16"
wide x 74
7
Is"
high. For the 6-foot unit, the
overall shipping dimensions are: 34
1 /32"
deep x 77
'/1s"
wide x 74
7
Is"
high.
Glass window
) /
(tron1 of oabine1)
Work surface opening
lower edge
Fig. 2 Adjusting the Telescoping Leg Assembly
6
~
0
J.l
\
CJ
0
vlegbolt
\\
-'-
,,
8"
lift
Fig. 3 Installation Without Lifting Equipment
2.
For convenience in moving the cabinet, its
depth
may
be reduced to 31
7
Is"
by removing
the light canopy assembly. First, with the
unit disconnected electrically, remove the two
nuts holding the canopy clips at the top of the
canopy. Then let the canopy pivot down
until the end restraints stop it. Disconnect
the electrical plugs inside on each end.
Disconnect the restraints and lift the canopy
up and off. 111en move the canopy to a new
location.
3.
The unit can be assembled with the work
surface either 30" or 36" from the floor
according to the needs of the owner. The unit
is shipped with the legs extending 6" below
t.lte cabirtet. If lifting
eqvipm-~nt
js available_.
lift the cabinet vertically to a height of
approximately 40." All four legs may then be
extended to the desired height by removing
the four shipping bolts, setting the height of
the legs and then securing the legs in position
with the four bolts which are supplied with
the unit
e
Is"
X
2
3
I
4" -
see Figure 2). use a
9 /16"
wrench. TI1e cabinet must be held safely
at the •±0" height duri.rtg the entire operation,
and set back onto the floor only after all four
legs have been secured. Final leveling may